Keysight (Agilent HP) 8594A Repair Service

8594A spectrum analyzer 9kHz-2.9GHz
Complete HP 8594A Spectrum Analyzer repair and calibration, even after factory support ends. Free estimates and quick turnaround times.
The Agilent Technologies 8594A spectrum analyzer is a wide band, very sensitive receiver. It works on the principle of "super-heterodyne receiver" to convert higher frequencies (normally ranging up to several 10s of GHz) to measurable quantities. The received frequency spectrum is slowly swept through a range of pre-selected frequencies, converting the selected frequency to a measurable DC level (usually logarithmic scale), and displaying the same on the CRT of the Agilent Technologies 8594A. The CRT displays received signal strength (y-axis) against frequency ( x-axis).
8594a
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
8594A Specifications – click to expand

8594A Specifications

Displayed average noise level:.

SpecificationValue
Frequency range9 kHz to 2.9 GHz
Span range0 Hz (zero span), 10 kHz to 2.9 GHz
Span accuracySpan 10 MHz: ±3%
Sweeptime RangeSpan = 0 Hz or > 10 kHz: 20 ms to 100 s
Sweeptime Accuracy20 ms to 100 s: ±3%
Amplitude range-112 to +30 dBm
8594A Options – click to expand

8594A Options

  • Options
  • 004: Precision Frequency Reference
  • 010: 50 ohm Built-in Tracking Generator
  • 021: HP-IB Interface
  • 023: RS-232 Interface
  • 101: Fast Time Domain Sweeps
  • 102: AM/FM Demodulator and TV Sync Trigger
  • 103: Quasi-Peak Detector/AM-FM Demodulator
  • 105: Time-gated Spectrum Analysis
  • 301: TV Sync Trigger
  • 1BH: general export version
HP 8594A Error Messages – click to expand

HP 8594A Error Messages

The HP 8594A portable spectrum analyzer (9 kHz to 2.9 GHz) shows text messages on screen when something is wrong. Each is marked (H) hardware, (U) user/setup, or (M) information.

First step: connect CAL OUT to the input and run CAL FREQ & AMPTD, then press CAL STORE. A hardware (H) message that comes back after a good self-cal means the 8594A needs repair.

8594A notes

  • The 8594A has no YIG-tuned filter, so CAL YTF and comb messages do not apply.
  • With Option 004 (precision frequency reference) the 8594A shows OVEN COLD for the first 5 minutes. That is normal.
  • The rear-panel 10 MHz REF OUT must be jumpered to EXT REF IN (or an external reference connected), or the 8594A shows REF UNLOCK.
  • Error codes and correction data are kept in battery-backed memory on A16. A dead BT101 battery shows up as default cal data or INVALID FLATNESS DATA.

8594A hardware messages

MessageTypeMeaningLikely assemblies
FAIL: xxxx xxxxxxxxxxHThe processor found a circuit fault during power-up or preset. The 4-digit and 10-digit code (hexadecimal) identifies the failed circuit.A7 analog interface, A16 processor/video, A17 memory card, A25 counter lock, card-cage boards
ADC-GND FAILHAt preset the A16 analog-to-digital converter ground reading is out of limits. ±15 V supplies or a board loading the bus can cause it.A7 analog interface, A8 power supply, A15 motherboard, A16 processor/video, card-cage boards
ADC-TIME FAILHTime between analog-to-digital readings is too long.A16 processor/video
ADC-2V FAILHThe A16 +2 V reference reading is out of limits.A8 power supply, A15 motherboard, A16 processor/video, card-cage boards
FREQ UNCALU/HThe YTO is more than 20 MHz from the frequency it should be at. Run CAL FREQ first; it can also flash briefly during CAL FREQ, which is normal.A3A7 YTO, A7 analog interface, A25 counter lock
CAL SIGNAL NOT FOUNDU/HCAL OUT was not found during self-cal (must be 300 MHz ±50 MHz and at least −45 dBm). Check the cable from CAL OUT to the input first.A7 analog interface, A9 third converter
CAL: GAIN FAILHSignal level at the ADC is too low during CAL AMPTD, even with the reference-level DAC at maximum.A3 front end, A7 analog interface, A9 third converter, A11/A13 bandwidth filters, A12 amplitude control, A14 log amp
CAL: SPAN SENS FAILHMain-coil span sensitivity failed (spans over 10 MHz). The 0 Hz, 300 MHz and 600 MHz signals were not found in a 750 MHz span.A7 analog interface, A3A7 YTO, A9 third converter, A16 processor/video
CAL: FM SPAN SENS FAILHThe analyzer could not set the FM coil span sensitivity.A3A7 YTO, A7 analog interface
CAL: RES BW AMPL FAILHA resolution bandwidth filter has more than 2.2 dB loss compared with the 3 kHz filter.A7 analog interface, A11/A13 bandwidth filters, A14 log amp/detector
CONF TEST FAILHThe confidence test failed. The message shown with it names the failed check.
POS-PK FAILHPositive-peak detector check failed during the confidence test.A16 processor/video
SAMPLE FAILHSample detector check failed during the confidence test.A16 processor/video
VID-BW FAILHVideo bandwidth check failed during the confidence test.A14 log amp/detector, A16 processor/video
RES-BW NOISE FAILHNoise floor did not drop when changing to the next narrower resolution bandwidth.A11/A13 bandwidth filters
RES-BW SHAPE FAILHA resolution bandwidth 3 dB width is not within 20% of nominal.A7 analog interface, A11/A13 bandwidth filters
STEP GAIN/ATTEN FAILHNoise did not rise with each 10 dB reference level step during the confidence test.A12 amplitude control
REF UNLOCKM/HThe A9 third converter oscillator is not locked to 10 MHz. Check the rear-panel jumper from 10 MHz REF OUT to EXT REF IN first.A22 precision reference (Option 004), A25 counter lock

8594A calibration, setup and information messages

MessageTypeMeaningLikely assemblies
CAL: USING DEFAULT DATAMCAL AMPTD did not finish, so default correction factors are in use.
CAL: DATA NOT STORED CAL AMP NEEDEDU/HDefault data is in use and cannot be stored. Run CAL FREQ & AMPTD, then CAL STORE.A16 processor/video
CAL: cannot execute CALAMP enter: 0 dB PREAMP GAINU/HSet EXT PREAMP gain to 0 dB before running CAL AMPTD.
CAL: PASSCODE NEEDEDMThe service function needs the passcode (center frequency −37 Hz).
INVALID FLATNESS DATAUFlatness correction constants are missing. Shows after a new A16 board, lost memory or a dead BT101 battery.A16 processor/video (BT101 battery)
OVEN COLDMOption 004 only: the precision reference oven is still warming up. Normal for 5 minutes after turn-on.A22 precision reference (Option 004)
SRQ ___MA service request is active. SRQ 110 usually comes with a hardware message.

Hardware message on your 8594A? Analyzer Repair Inc. repairs and calibrates the HP 8594A in-house in Cocoa, Florida, even though HP support ended long ago. Tell us the message you see; estimates are free once the unit is in our lab. Contact us for service.

Message list based on the HP 8595A Service Manual (08595-90001), Chapter 11 "Analyzer Messages".

HP 8594A Replaceable Parts – click to expand

HP 8594A Replaceable Assemblies and Part Numbers

These are the major replaceable assemblies in the HP 8594A, with their HP / Agilent part numbers. Option-specific assemblies are only present if your 8594A has that option installed. "Exchange" numbers are rebuilt units traded for your old assembly.

ReferenceAssemblyHP part numberOption
A1Front Frame08594-60004
A1Front Frame08594-60009103
A1A1Keyboard Board08590-60004
A1A2Rotary Pulse Generator0960-0745
A1R1Volume-Intensity Dual Pot08590-60133
A2CRT Display (cables and shield not included)5062-6404
A3A3Low-Pass Filter 2.9 GHz0955-0420
A3A4Second Converter5086-7819
A3A5Attenuator 0 to 70 dB / DC Block5086-7777
A3A6Low Band Mixer5086-7748
A3A7YIG-Tuned Oscillator (YTO) 3.0 to 6.82 GHz5086-7833
A3A9Bandpass Filter, 2nd IF9135-0252
A3A10Directional Coupler0955-0098
A3A13Isolator 6.6 GHz0955-0204
A3A14First LO Distribution Amplifier5086-7744009, 010
A3A15Tracking Generator5086-7840010
A7Analog Interface Board08594-60001
A8Power Supply 100 W0950-1977
A9Third Converter Board08590-60192
A9Third Converter Board08593-60021010
A10Tracking Generator Control Board5062-6432010
A10LO Distribution Amplifier Control Board5062-8219009
A11, A13Bandwidth Filter Boards #1 and #208590-60195
A12Amplitude Control Board08590-60105
A14Log Amp Board08590-60149
A15Motherboard08591-60017
A16Processor/Video Board08593-60005
A17Memory Card Board08590-60107003
A20HP-IB Interface Board08590-60108021
A21RS-232 Interface Board08590-60109023
A2210 MHz Oven Crystal Oscillator (OCXO)1813-0644004
A25Counter Lock Board08591-60027
A101Fast Time Domain Sweep Board5062-7079101
A102AM/FM Speaker and TV Sync Trigger Board5062-1982102
A103Quasi-Peak Detector Board5062-1983103
A105Time Gate Board5062-7754105
B1Fan Assembly08590-60166
BT101Battery 3.6 V 1.7 Ah1420-0331

Part obsolete or unavailable? Many 8594A assemblies are no longer sold by Keysight. Analyzer Repair Inc. does not sell parts; we repair the 8594A, including these boards and modules, at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

Part numbers from the HP 8594A Preliminary Service Information (in HP 8591A Service Manual 08591-90018), Table 4-2 "HP 8594A Assembly-Level Replaceable Parts". They are listed for reference and identification only.

Located in: Spectrum Analyzers