Keysight (Agilent HP) 8719ET Repair Service

8719ET Trans/Reflection Network Analyzer, 13.5 GHz
Keep your Agilent HP 8719ET Network Analyzer in service with our repair and calibration. We specialize in unsupported equipment. Free evaluations.
We are specialists in the repair of the Agilent/HP 8719ET Network Analyzers. Keysight Agilent HP 8719ET vector network analyzers perform complete characterization of RF components.
8719et
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
8719ET Specifications – click to expand

8719ET Specifications

GPIB interface.

SpecificationValue
System Trace noise, Magnitude (RMS of peak-to-peak value)0.05 to 13.5 GHz <0.03 dB, 13.5 to 20 GHz <0.04 dB, 20 to 40 GHz <0.15 dB
System Trace noise, Phase (RMS of peak-to-peak value)0.05 to 13.5 GHz <0.3°, 13.5 to 20 GHz <0.4°, 20 to 40 GHz <1.5°
Spectral purity, Harmonics at maximum output level<–15 dBc
Spectral purity, Phase noiseto 60 kHz from carrier at 2 GHz <–55 dBc, to 60 kHz from carrier at 20 GHz <–35 dBc
Spectral purity, Nonharmonic spurious signalsat 100 kHz offset <–40 dBc, at 200 kHz offset <–45 dBc, at >200 kHz offset <–65 dBc
Power range20 dB
Resolution Level accuracy (at 5 dB below maximum output power)±2 dB
Power sweep range20 dB (30 dB typical)
Power linearity–5 dB from reference, ±0.35; +5 dB from reference, ±0.35; –10 dB from reference, ±0.60; +10 dB from reference, ±1.0
System Dynamic range (standard, Options 004)0.05 to 0.84 GHz, 102 dB; 0.84 to 20 GHz, 104 dB
Measured number of points per sweepLinear frequency: choose 3, 11, 21, 51, 101, 201, 401, 801, or 1601 points
8719ET Options – click to expand

8719ET Options

  • 004, Step Attenuator
  • 010, Time Domain
  • 1D5, High-Stability Frequency Reference
HP / Agilent 8719ET Error Messages – click to expand

HP / Agilent 8719ET Error Messages

The Agilent 8719ET (50 MHz to 13.5 GHz) reports faults as text messages with an error number. Hardware (H) messages point to a failed assembly; user (U) messages mean a setup, calibration or cable issue; (M) messages are information.

8719ET notes

  • The 8719ET is the transmission/reflection version: it measures forward only and has no transfer switch, so test-port switching and attenuator messages do not apply.
  • Phase lock errors on the 8719ET usually point to the source (YIG oscillator or source assembly) or the phase lock board, but check R channel input level first.

8719ET hardware error messages

MessageError no.TypeMeaning
PHASE LOCK FAILURE7HThe analyzer could not acquire phase lock. Check R channel level, then the source and phase lock circuits.
PHASE LOCK LOST8HPhase lock was acquired and then dropped during the sweep.
PHASE LOCK CAL FAILED4HThe automatic phase lock calibration (run at power-on or when pretune drifts) failed.
POSSIBLE FALSE LOCK6HThe source may be locked to the wrong harmonic comb tooth, so frequencies can be wrong.
SWEEP TIME TOO FAST12HThe fractional-N and digital IF circuits lost synchronization; points to the digital control section.
CAUTION: POWER OUT MAY BE UNLEVELED179HSource output may not be leveled. Can be a power setting beyond the source range or a source/ALC fault.
POWER SUPPLY HOT!21HThe power supply is overheating. Check the fan, filter and air flow.
POWER SUPPLY SHUT DOWN!22HOne or more supplies shut down, usually from an overload or short on a board.
BATTERY FAILED. STATE MEMORY CLEARED183HThe memory backup battery has failed and stored instrument states were cleared. Replace the battery on the CPU board.
BATTERY LOW! STORE SAVE REGS TO DISK184HThe memory backup battery is getting weak. Save your states to disk and replace the battery.
CORRECTION CONSTANTS NOT STORED3HWriting correction constants to EEPROM failed. The A9 CPU write-protect switch/jumper must be set to allow writing.
DEADLOCK111HAn internal firmware deadlock occurred. Cycle power; if it repeats, the CPU or a firmware problem is likely.
DISK HARDWARE PROBLEM39HThe disk drive is not responding properly.
DISK READ/WRITE ERROR189HA disk read or write failed; try another disk, then suspect the drive.
DISK MESSAGE LENGTH ERROR190HThe disk drive and analyzer are out of step during a transfer.
INITIALIZATION FAILED47HThe disk could not be initialized; check the disk and drive.
NO IF FOUND: CHECK R INPUT LEVEL5HThe first IF was not detected while pretuning the phase lock. Usually a source fault, or the R channel is getting too little signal (needs at least −35 dBm).

8719ET calibration, setup and information messages

MessageError no.TypeMeaning
SOURCE POWER TURNED OFF, RESET UNDER POWER MENUMAn input overload tripped the source off. Remove the overload and turn power back on in the power menu.
CALIBRATION REQUIRED63UNo calibration set matches the current stimulus or measurement; perform a new cal.
CURRENT PARAMETER NOT IN CAL SET64UThe parameter you selected was not included in the active calibration.
CORRECTION TURNED OFF66USweep type, start, span or number of points changed from the cal, so correction was switched off.
NO CALIBRATION CURRENTLY IN PROGRESS69UA cal step was requested but no calibration is running.
NO SPACE FOR NEW CAL. CLEAR REGISTERS70UMemory is full; clear saved registers to store a new calibration.
CALIBRATION ABORTED74UThe active channel was changed during a cal, so it stopped. Restart on the right channel.
POWER METER INVALID116UThe power meter reading is out of range or not valid.
PWR MTR: NOT ON/CONNECTED OR WRONG ADDRS117UThe power meter is off, not connected by GPIB, or set to the wrong address.
POWER METER NOT SETTLED118UPower meter readings are not stable.
NOT ALLOWED DURING POWER METER CAL198UThat function cannot be used while a source power meter calibration is running.
INSUFFICIENT MEMORY, PWR MTR CAL OFF154UNot enough memory for power meter cal data, so it was turned off.
SAVE FAILED. INSUFFICIENT MEMORY151UNot enough memory to save the state; clear some registers.
NOT ENOUGH SPACE ON DISK FOR STORE44UThe disk is full.
NO FILE(S) FOUND ON DISK45UNo files of the requested type are on the disk.
PARALLEL PORT NOT AVAILABLE FOR COPY167UThe parallel port is set for GPIO, so it cannot be used for printing.
PARALLEL PORT NOT AVAILABLE FOR GPIO165UThe parallel port is set for printing, so it cannot be used as GPIO.
SWEEP MODE CHANGED TO CW TIME SWEEP187MThe sweep mode was switched to CW time sweep.
OVERLOAD ON INPUT A, POWER REDUCED58UToo much power at input A; the analyzer reduced source power (P⇓ annotation). Lower the power, then reset it in the power menu.
OVERLOAD ON INPUT B, POWER REDUCED59UToo much power at input B; source power was reduced (P⇓). Lower the power, then reset it in the power menu.
TEST PORT OVERLOAD, REDUCE POWER57UToo much power at a test port; source power was reduced. Lower the input level.
ADDITIONAL STANDARDS NEEDED68UNot all calibration standards for this cal type have been measured yet.
ANALOG INPUT OVERLOAD60UThe signal on the analog (AUX) input is too high.
ILLEGAL UNIT OR VOLUME NUMBER46UThe disk unit or volume number is not valid.
CANNOT READ/WRITE 1 FILE SYSTEM203UThe disk uses a file system the analyzer cannot read or write.
ADDRESSED TO TALK WITH NOTHING TO SAY31UA GPIB read was sent with no data waiting in the output queue.
BLOCK INPUT ERROR34UThe analyzer received bad data during a GPIB block transfer.
BLOCK INPUT LENGTH ERROR35UA GPIB block transfer had the wrong length.
ANALOG BUS DISABLED IN 6KHZ IF BW212MWith the 6 kHz IF bandwidth selected, the service analog bus is not available.

Hardware error on your 8719ET? Analyzer Repair Inc. repairs and calibrates the HP / Agilent 8719ET in-house in Cocoa, Florida, including source, phase lock, sampler and test set faults. Free estimate once the unit is in our lab. Contact us for service.

Message list based on the Agilent 8719ET/ES, 8720ET/ES, 8722ET/ES Service Guide (08720-90397), service error messages. Measurement-only messages are in the 8719ET reference guide.

HP / Agilent 8719ET Replaceable Parts – click to expand

HP 8719ET Replaceable Assemblies and Part Numbers

Major replaceable assemblies in the 8719ET, with HP / Agilent part numbers from the Agilent 8719ET/ES, 8720ET/ES, 8722ET/ES Service Guide (08720-90397). Check your option list before ordering option-specific parts.

ReferenceAssemblyHP part numberNotes
A1Keyboard board08720-60267
A2Front panel interface board08720-60260
A3Disk drive replacement kit08720-60190
A4, A5, A6Second converter08720-60156
A7CPU repair kit08720-60253Exchange 08720-69253
A8Post regulator board08722-60099Exchange 08722-69099
A9Source control board08720-60129
A10Digital IF board08753-60958
A11Phase lock board08720-60181
A12Reference board08720-60252Exchange 08720-69252
A13Frac-N analog board08720-60258Exchange 08720-69258
A14Frac-N digital board08720-60179Exchange 08720-69179
A15Preregulator08753-60098Exchange 08753-69098
A17Motherboard08720-60264
A18Color LCD assembly2090-0361
A19GSP (graphics) board08720-60261
A20, A29Microwave switch08415-60057
A28Blocking capacitor11742A
A52Pulse generator5086-7456Exchange 5086-6456
A53Low band assembly5086-7583Exchange 5086-6583
A55YIG oscillator, 2.4 to 13.5 GHz5087-7081
A57Fixed oscillator08720-60073
A58M/A/D/S (modulator/amplifier/detector/splitter)5087-7119Standard
A59Source interface board08720-60322
A64, A65, A66Sampler5086-7614Exchange 5086-6614
A68Attenuator, 6 dB0955-0462
A98Frequency converter08720-60121Exchange 08720-69121

Part obsolete? Many 8719ET assemblies are no longer sold new. Analyzer Repair Inc. repairs 8719ET boards, samplers and RF modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

From the Agilent 8719ET/ES, 8720ET/ES, 8722ET/ES Service Guide (08720-90397).