Keysight (Agilent HP) 8719ET Repair Service
Product model
8719ET Trans/Reflection Network Analyzer, 13.5 GHzServices
Description

Repair prices
8719ET Specifications – click to expand
8719ET Specifications
GPIB interface.
| Specification | Value |
|---|---|
| System Trace noise, Magnitude (RMS of peak-to-peak value) | 0.05 to 13.5 GHz <0.03 dB, 13.5 to 20 GHz <0.04 dB, 20 to 40 GHz <0.15 dB |
| System Trace noise, Phase (RMS of peak-to-peak value) | 0.05 to 13.5 GHz <0.3°, 13.5 to 20 GHz <0.4°, 20 to 40 GHz <1.5° |
| Spectral purity, Harmonics at maximum output level | <–15 dBc |
| Spectral purity, Phase noise | to 60 kHz from carrier at 2 GHz <–55 dBc, to 60 kHz from carrier at 20 GHz <–35 dBc |
| Spectral purity, Nonharmonic spurious signals | at 100 kHz offset <–40 dBc, at 200 kHz offset <–45 dBc, at >200 kHz offset <–65 dBc |
| Power range | 20 dB |
| Resolution Level accuracy (at 5 dB below maximum output power) | ±2 dB |
| Power sweep range | 20 dB (30 dB typical) |
| Power linearity | –5 dB from reference, ±0.35; +5 dB from reference, ±0.35; –10 dB from reference, ±0.60; +10 dB from reference, ±1.0 |
| System Dynamic range (standard, Options 004) | 0.05 to 0.84 GHz, 102 dB; 0.84 to 20 GHz, 104 dB |
| Measured number of points per sweep | Linear frequency: choose 3, 11, 21, 51, 101, 201, 401, 801, or 1601 points |
8719ET Options – click to expand
8719ET Options
- 004, Step Attenuator
- 010, Time Domain
- 1D5, High-Stability Frequency Reference
HP / Agilent 8719ET Error Messages – click to expand
HP / Agilent 8719ET Error Messages
The Agilent 8719ET (50 MHz to 13.5 GHz) reports faults as text messages with an error number. Hardware (H) messages point to a failed assembly; user (U) messages mean a setup, calibration or cable issue; (M) messages are information.
8719ET notes
- The 8719ET is the transmission/reflection version: it measures forward only and has no transfer switch, so test-port switching and attenuator messages do not apply.
- Phase lock errors on the 8719ET usually point to the source (YIG oscillator or source assembly) or the phase lock board, but check R channel input level first.
8719ET hardware error messages
| Message | Error no. | Type | Meaning |
|---|---|---|---|
| PHASE LOCK FAILURE | 7 | H | The analyzer could not acquire phase lock. Check R channel level, then the source and phase lock circuits. |
| PHASE LOCK LOST | 8 | H | Phase lock was acquired and then dropped during the sweep. |
| PHASE LOCK CAL FAILED | 4 | H | The automatic phase lock calibration (run at power-on or when pretune drifts) failed. |
| POSSIBLE FALSE LOCK | 6 | H | The source may be locked to the wrong harmonic comb tooth, so frequencies can be wrong. |
| SWEEP TIME TOO FAST | 12 | H | The fractional-N and digital IF circuits lost synchronization; points to the digital control section. |
| CAUTION: POWER OUT MAY BE UNLEVELED | 179 | H | Source output may not be leveled. Can be a power setting beyond the source range or a source/ALC fault. |
| POWER SUPPLY HOT! | 21 | H | The power supply is overheating. Check the fan, filter and air flow. |
| POWER SUPPLY SHUT DOWN! | 22 | H | One or more supplies shut down, usually from an overload or short on a board. |
| BATTERY FAILED. STATE MEMORY CLEARED | 183 | H | The memory backup battery has failed and stored instrument states were cleared. Replace the battery on the CPU board. |
| BATTERY LOW! STORE SAVE REGS TO DISK | 184 | H | The memory backup battery is getting weak. Save your states to disk and replace the battery. |
| CORRECTION CONSTANTS NOT STORED | 3 | H | Writing correction constants to EEPROM failed. The A9 CPU write-protect switch/jumper must be set to allow writing. |
| DEADLOCK | 111 | H | An internal firmware deadlock occurred. Cycle power; if it repeats, the CPU or a firmware problem is likely. |
| DISK HARDWARE PROBLEM | 39 | H | The disk drive is not responding properly. |
| DISK READ/WRITE ERROR | 189 | H | A disk read or write failed; try another disk, then suspect the drive. |
| DISK MESSAGE LENGTH ERROR | 190 | H | The disk drive and analyzer are out of step during a transfer. |
| INITIALIZATION FAILED | 47 | H | The disk could not be initialized; check the disk and drive. |
| NO IF FOUND: CHECK R INPUT LEVEL | 5 | H | The first IF was not detected while pretuning the phase lock. Usually a source fault, or the R channel is getting too little signal (needs at least −35 dBm). |
8719ET calibration, setup and information messages
| Message | Error no. | Type | Meaning |
|---|---|---|---|
| SOURCE POWER TURNED OFF, RESET UNDER POWER MENU | M | An input overload tripped the source off. Remove the overload and turn power back on in the power menu. | |
| CALIBRATION REQUIRED | 63 | U | No calibration set matches the current stimulus or measurement; perform a new cal. |
| CURRENT PARAMETER NOT IN CAL SET | 64 | U | The parameter you selected was not included in the active calibration. |
| CORRECTION TURNED OFF | 66 | U | Sweep type, start, span or number of points changed from the cal, so correction was switched off. |
| NO CALIBRATION CURRENTLY IN PROGRESS | 69 | U | A cal step was requested but no calibration is running. |
| NO SPACE FOR NEW CAL. CLEAR REGISTERS | 70 | U | Memory is full; clear saved registers to store a new calibration. |
| CALIBRATION ABORTED | 74 | U | The active channel was changed during a cal, so it stopped. Restart on the right channel. |
| POWER METER INVALID | 116 | U | The power meter reading is out of range or not valid. |
| PWR MTR: NOT ON/CONNECTED OR WRONG ADDRS | 117 | U | The power meter is off, not connected by GPIB, or set to the wrong address. |
| POWER METER NOT SETTLED | 118 | U | Power meter readings are not stable. |
| NOT ALLOWED DURING POWER METER CAL | 198 | U | That function cannot be used while a source power meter calibration is running. |
| INSUFFICIENT MEMORY, PWR MTR CAL OFF | 154 | U | Not enough memory for power meter cal data, so it was turned off. |
| SAVE FAILED. INSUFFICIENT MEMORY | 151 | U | Not enough memory to save the state; clear some registers. |
| NOT ENOUGH SPACE ON DISK FOR STORE | 44 | U | The disk is full. |
| NO FILE(S) FOUND ON DISK | 45 | U | No files of the requested type are on the disk. |
| PARALLEL PORT NOT AVAILABLE FOR COPY | 167 | U | The parallel port is set for GPIO, so it cannot be used for printing. |
| PARALLEL PORT NOT AVAILABLE FOR GPIO | 165 | U | The parallel port is set for printing, so it cannot be used as GPIO. |
| SWEEP MODE CHANGED TO CW TIME SWEEP | 187 | M | The sweep mode was switched to CW time sweep. |
| OVERLOAD ON INPUT A, POWER REDUCED | 58 | U | Too much power at input A; the analyzer reduced source power (P⇓ annotation). Lower the power, then reset it in the power menu. |
| OVERLOAD ON INPUT B, POWER REDUCED | 59 | U | Too much power at input B; source power was reduced (P⇓). Lower the power, then reset it in the power menu. |
| TEST PORT OVERLOAD, REDUCE POWER | 57 | U | Too much power at a test port; source power was reduced. Lower the input level. |
| ADDITIONAL STANDARDS NEEDED | 68 | U | Not all calibration standards for this cal type have been measured yet. |
| ANALOG INPUT OVERLOAD | 60 | U | The signal on the analog (AUX) input is too high. |
| ILLEGAL UNIT OR VOLUME NUMBER | 46 | U | The disk unit or volume number is not valid. |
| CANNOT READ/WRITE 1 FILE SYSTEM | 203 | U | The disk uses a file system the analyzer cannot read or write. |
| ADDRESSED TO TALK WITH NOTHING TO SAY | 31 | U | A GPIB read was sent with no data waiting in the output queue. |
| BLOCK INPUT ERROR | 34 | U | The analyzer received bad data during a GPIB block transfer. |
| BLOCK INPUT LENGTH ERROR | 35 | U | A GPIB block transfer had the wrong length. |
| ANALOG BUS DISABLED IN 6KHZ IF BW | 212 | M | With the 6 kHz IF bandwidth selected, the service analog bus is not available. |
Hardware error on your 8719ET? Analyzer Repair Inc. repairs and calibrates the HP / Agilent 8719ET in-house in Cocoa, Florida, including source, phase lock, sampler and test set faults. Free estimate once the unit is in our lab. Contact us for service.
Message list based on the Agilent 8719ET/ES, 8720ET/ES, 8722ET/ES Service Guide (08720-90397), service error messages. Measurement-only messages are in the 8719ET reference guide.
HP / Agilent 8719ET Replaceable Parts – click to expand
HP 8719ET Replaceable Assemblies and Part Numbers
Major replaceable assemblies in the 8719ET, with HP / Agilent part numbers from the Agilent 8719ET/ES, 8720ET/ES, 8722ET/ES Service Guide (08720-90397). Check your option list before ordering option-specific parts.
| Reference | Assembly | HP part number | Notes |
|---|---|---|---|
| A1 | Keyboard board | 08720-60267 | |
| A2 | Front panel interface board | 08720-60260 | |
| A3 | Disk drive replacement kit | 08720-60190 | |
| A4, A5, A6 | Second converter | 08720-60156 | |
| A7 | CPU repair kit | 08720-60253 | Exchange 08720-69253 |
| A8 | Post regulator board | 08722-60099 | Exchange 08722-69099 |
| A9 | Source control board | 08720-60129 | |
| A10 | Digital IF board | 08753-60958 | |
| A11 | Phase lock board | 08720-60181 | |
| A12 | Reference board | 08720-60252 | Exchange 08720-69252 |
| A13 | Frac-N analog board | 08720-60258 | Exchange 08720-69258 |
| A14 | Frac-N digital board | 08720-60179 | Exchange 08720-69179 |
| A15 | Preregulator | 08753-60098 | Exchange 08753-69098 |
| A17 | Motherboard | 08720-60264 | |
| A18 | Color LCD assembly | 2090-0361 | |
| A19 | GSP (graphics) board | 08720-60261 | |
| A20, A29 | Microwave switch | 08415-60057 | |
| A28 | Blocking capacitor | 11742A | |
| A52 | Pulse generator | 5086-7456 | Exchange 5086-6456 |
| A53 | Low band assembly | 5086-7583 | Exchange 5086-6583 |
| A55 | YIG oscillator, 2.4 to 13.5 GHz | 5087-7081 | |
| A57 | Fixed oscillator | 08720-60073 | |
| A58 | M/A/D/S (modulator/amplifier/detector/splitter) | 5087-7119 | Standard |
| A59 | Source interface board | 08720-60322 | |
| A64, A65, A66 | Sampler | 5086-7614 | Exchange 5086-6614 |
| A68 | Attenuator, 6 dB | 0955-0462 | |
| A98 | Frequency converter | 08720-60121 | Exchange 08720-69121 |
Part obsolete? Many 8719ET assemblies are no longer sold new. Analyzer Repair Inc. repairs 8719ET boards, samplers and RF modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.
From the Agilent 8719ET/ES, 8720ET/ES, 8722ET/ES Service Guide (08720-90397).