Keysight (Agilent HP) 8590L Repair Service

8590L spectrum analyzer 9kHz-1.8GHz
Keep your Agilent HP 8590L Spectrum Analyzer in service with our repair and calibration. We specialize in unsupported equipment. Free evaluations.
The Agilent 8590L is a low cost, but full-featured, frequency accurate RF spectrum analyzer designed to meet your general purpose measurement needs. The easy-to-use interface provides access to more than 200 built-in functions.
8590l
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
8590L Specifications – click to expand

8590L Specifications

  • Portable spectrum analyzer
  • Resolution bandwidths to 1 kHz (30 Hz with Option 130 on E-series)
  • HP-IB, RS-232 and parallel interfaces available by option
SpecificationValue
Frequency range9 kHz to 1.8 GHz
8590L Options – click to expand

8590L Options

  • 004, Precision Frequency Reference
  • 010, Built-In Tracking Generator
  • 021, HP-IB Interface
  • 023, RS-232 Interface
  • 041, HP-IB and Parallel Interface
  • 043, RS-232 and Parallel Interface
  • 101, Fast Time-Domain Sweeps
  • 105, Time-Gated Spectrum Analysis
  • 130, Narrow Resolution Bandwidths
HP / Agilent 8590L Error Messages – click to expand

HP / Agilent 8590L Error Messages

When the HP / Agilent 8590L (9 kHz to 1.8 GHz) detects a problem, it shows a text message on screen rather than a numbered code. Messages are marked (H) hardware, (U) user/setup, or (M) information.

Try this first: connect CAL OUT to the input with the supplied cable and run CAL FREQ & AMPTD, then press CAL STORE. Many messages clear after a good self-cal. If an (H) message keeps returning, the 8590L needs repair.

8590L notes

  • The 8590L uses the A6 YTO and has no YIG-tuned filter, so CAL YTF and preselector messages do not apply.
  • Memory card functions on the 8590L need Option 003; without it, card commands give INVALID CARD messages.
  • Tracking generator options for the 8590L: Option 010 (50 ohm, 100 kHz to 1.8 GHz) or 011 (75 ohm, 1 MHz to 1.8 GHz).
  • The rear-panel 10 MHz REF OUT must be jumpered to EXT REF IN (or an external reference connected), or the 8590L shows REF UNLOCK.

Most common 8590L hardware messages

MessageTypeMeaningLikely assemblies
FREQ UNCALU/HYTO is more than 20 MHz off the frequency it should be on. Run CAL FREQ and CAL STORE; if it stays, suspect the YTO loop or an unlocked reference/sampler on A25.A6 YTO, A7 analog interface, A25 counter lock
CAL SIGNAL NOT FOUNDU/HSelf-cal could not find the 300 MHz CAL OUT signal (must be within 50 MHz of 300 MHz and above −45 dBm). Check the CAL cable first. With default cal data loaded, the −37 Hz passcode bypasses this check.A7 analog interface, A9 third converter
ADC-2V FAILHAt preset the ADC reading of the +2 V reference on the A16 processor/video board is out of limits. Often caused by a bad ±15 V supply or a card-cage board pulling an A16J1 line high.A8 power supply, A15 motherboard, A16 processor/video, card-cage boards
ADC-GND FAILHAt preset the ADC ground reading on A16 is out of limits. Same usual suspects as ADC-2V FAIL: ±15 V supplies or a board loading the A16J1 pins.A7 analog interface, A8 power supply, A15 motherboard, A16 processor/video, card-cage boards
REF UNLOCKM/HThe A9 third converter oscillator is not locked to 10 MHz. Check the rear-panel jumper from 10 MHz REF OUT to EXT REF IN first, then the reference level and cables.A22 precision reference (Option 004), A25 counter lock, W17/W18 cables
SMPLR UNLCKU/HSampling oscillator is not working correctly. Check the external reference connection, then the frequency diagnostics.A6 YTO, A25 counter lock
CAL: GAIN FAILHDuring CAL AMPTD the signal at the ADC stays too low even with the reference-level DAC at maximum. Trace the signal from the RF input through the IF to the log amp.A3 input attenuator, A4 first converter, A5 second converter, A6 YTO, A7, A9, A11, A12, A13, A14
CAL: SPAN SENS FAILHMain-coil span calibration failed: it could not find the 0 Hz, 300 MHz and 600 MHz signals in a 750 MHz span. Spurious signals can also cause it.A6 YTO, A7, A9, A16
USING DEFAULTS: _ _ _HSome memory locations failed their check at power-on. Often a dead memory backup battery.A16 processor/video (battery)
FAIL: xxxx xxxxxxxxxxHPower-up circuit test failure on the A16 processor board. See the FAIL code section below.A16, A7, A17 memory card, A25, card-cage boards

Complete 8590L hardware and calibration message list

MessageTypeMeaningLikely assemblies
ADC-2V FAILHAt preset the ADC reading of the +2 V reference on the A16 processor/video board is out of limits. Often caused by a bad ±15 V supply or a card-cage board pulling an A16J1 line high.A8 power supply, A15 motherboard, A16 processor/video, card-cage boards
ADC-GND FAILHAt preset the ADC ground reading on A16 is out of limits. Same usual suspects as ADC-2V FAIL: ±15 V supplies or a board loading the A16J1 pins.A7 analog interface, A8 power supply, A15 motherboard, A16 processor/video, card-cage boards
ADC-TIME FAILHTime between ADC readings is too long. If the unit will not sweep but everything else works, A16 is the likely cause.A16 processor/video
φ LOCK OFFU/HYTO is tuning slowly. Normal for a moment during CAL FREQ, preset or a frequency change. If it stays on, run CAL FREQ.A6 YTO, A9 third converter, A25 counter lock
FREQ UNCALU/HYTO is more than 20 MHz off the frequency it should be on. Run CAL FREQ and CAL STORE; if it stays, suspect the YTO loop or an unlocked reference/sampler on A25.A6 YTO, A7 analog interface, A25 counter lock
REF UNLOCKM/HThe A9 third converter oscillator is not locked to 10 MHz. Check the rear-panel jumper from 10 MHz REF OUT to EXT REF IN first, then the reference level and cables.A22 precision reference (Option 004), A25 counter lock, W17/W18 cables
SMPLR UNLCKU/HSampling oscillator is not working correctly. Check the external reference connection, then the frequency diagnostics.A6 YTO, A25 counter lock
CAL SIGNAL NOT FOUNDU/HSelf-cal could not find the 300 MHz CAL OUT signal (must be within 50 MHz of 300 MHz and above −45 dBm). Check the CAL cable first. With default cal data loaded, the −37 Hz passcode bypasses this check.A7 analog interface, A9 third converter
CAL: GAIN FAILHDuring CAL AMPTD the signal at the ADC stays too low even with the reference-level DAC at maximum. Trace the signal from the RF input through the IF to the log amp.A3 input attenuator, A4 first converter, A5 second converter, A6 YTO, A7, A9, A11, A12, A13, A14
CAL: SPAN SENS FAILHMain-coil span calibration failed: it could not find the 0 Hz, 300 MHz and 600 MHz signals in a 750 MHz span. Spurious signals can also cause it.A6 YTO, A7, A9, A16
CAL: FM SPAN SENS FAILHCould not set the FM coil span sensitivity during CAL FREQ. Run CAL FREQ again.A6 YTO, A7 analog interface
CAL: MAIN COIL SENSE FAILHCould not set the main coil span sensitivity. Load DEFAULT CAL DATA (−37 Hz passcode) and run CAL FREQ again.A6 YTO, A7 analog interface
CAL: ZERO FAILHCould not set the main coil tuning sensitivity. Load DEFAULT CAL DATA and run CAL FREQ again.A6 YTO, A7 analog interface
CAL: RES BW AMPL FAILHDuring CAL AMPTD a resolution bandwidth filter had more than 2.2 dB loss compared to the 3 kHz filter. The routine stops on the failed bandwidth.A7, A11/A13 bandwidth filters, A14 log amp
CAL: NBW gain / width / 200 Hz notch failedHA narrow resolution bandwidth (including 200 Hz) has the wrong gain, width or shape during CAL AMPTD.A130 narrow RBW board
CAL: DATA NOT STORED CAL AMP NEEDEDU/HYou tried to store default cal data. Run CAL FREQ and CAL AMPTD first, then CAL STORE.A16 processor/video
CAL: USING DEFAULT DATAMCAL AMPTD did not finish, so default correction factors are in use.
USING DEFAULTS self cal neededUDefault correction factors are loaded. Run CAL FREQ & AMPTD.
USING DEFAULTS: _ _ _HSome memory locations failed their check at power-on. Often a dead memory backup battery.A16 processor/video (battery)
CAL: cannot execute CALAMP enter: 0 dB PREAMP GAINU/HExternal preamp gain must be 0 dB before CAL AMPTD. If it is already 0 dB, suspect A16 RAM.A16 processor/video
YTF is not availableUThe 8590L has no YIG-tuned filter, so YTF functions do not apply.
CONF TEST FAILU/HThe confidence test failed; another message names the failed check. Connect CAL OUT, run CAL FREQ & AMPTD, and test again.
POS-PK FAILHConfidence test: the positive-peak detector failed.A16 processor/video
SAMPLE FAILHConfidence test: the sample detector failed.A16 processor/video
VID-BW FAILHConfidence test: the video bandwidths are not reducing the noise as they should.A14 log amp/detector, A16
RES-BW NOISE FAILHConfidence test: the noise floor did not drop when stepping to a narrower resolution bandwidth. Loose IF cover screws can cause it.A11/A13 bandwidth filters, A7 control lines
RES-BW SHAPE FAILHConfidence test: a resolution bandwidth (10 kHz to 3 MHz) is more than 20% off its nominal 3 dB width.A7, A11/A13 bandwidth filters
STEP GAIN/ATTEN FAILHConfidence test: the 10 dB IF step gains did not raise the noise with each reference level step.A12 amplitude control, A7 control lines
FAIL: xxxx xxxxxxxxxxHPower-up circuit test failure on the A16 processor board. See the FAIL code section below.A16, A7, A17 memory card, A25, card-cage boards
INVALID SCRMOVEHPossible hardware failure. Remove the unit from the bus and preset.Any internal assembly
SYMTAB EMPTY / EMPTY DLP MEMUUser-defined functions and personalities in memory were deleted. If it shows at every power-up, suspect hardware.A16 processor/video
Memory Overflow: ERASE DLP MEM and reloadUToo many downloadable programs or user items loaded. Use DISPOSE USER MEM, then reload.A16 processor/video
OVEN COLDMShown for 5 minutes after power-on on units with the Option 004 precision reference. It is a timer only; the oven temperature is not measured.A22 precision frequency reference
TG UNLVLU/HTracking generator unleveled: power set out of range, span beyond the generator range, or bad TG cal data.A7A1, A24 tracking generator
TG SIGNAL NOT FOUNDUDuring CAL TRK GEN no signal was seen; check the cable from RF OUT to the input.A7A1, A24 tracking generator
Gate card not calibratedU/HTime gating (Option 105) needs CAL AMPTD with nothing connected to GATE TRIGGER INPUT.A105 gated video board
Cannot engage phase lock with current CAL FREQ dataURun CAL FREQ before turning on phase lock.A9 third converter
MEAS UNCALUSweep time, span and bandwidth settings make the measurement uncalibrated. Press AUTO COUPLE, AUTO ALL.
CAL: PASSCODE NEEDED / PASSCODE NEEDEDM/UThe function needs the service passcode (center frequency −37 Hz).

8590L power-up FAIL codes

A message like FAIL: DF40 0625C1F72D means the 8590L processor (A16) found a circuit fault at power-up or preset. The first 4 hex digits match the 16 LEDs on the edge of A16 and point to the failed circuit. The 10-digit part is three codes: I/O address bus (2 digits), I/O data bus (4 digits) and A16 video RAM (4 digits). Convert each hex digit to binary to find the failed line.

First 4 digits (bit)A16 LEDCircuitA16 part
1st digit, 816Video RAMU305, U306
1st digit, 415I/O bus addressU18
1st digit, 214CPUU12
1st digit, 11368230 timer/portU57
2nd digit, 812I/O bus (odd byte)U2
2nd digit, 411I/O bus (even byte)U3
2nd digit, 2 / 3rd digit, 810 / 8User RAMU5
2nd digit, 1 / 3rd digit, 49 / 7User RAMU22
4th digit, 84ROM (odd B)U6
4th digit, 43ROM (even B)U23
4th digit, 22ROM (odd A)U7
4th digit, 11ROM (even A)U24

A video RAM code means the A16 board itself is bad. For I/O bus address or data codes, power down and remove the related boards one at a time; if the code clears, the last board removed is the fault, otherwise suspect A16.

8590L user and setup messages (U)

These mean a setting, cable, card or remote command problem, not a hardware fault.

AreaMessages
Memory cardINVALID CARD, INVALID CARD: BAD MEDIA / DATA ERROR / DIRECTORY / NO CARD / TYPE, No card found, INVALID FILE: NO ROOM, File type incompatible
Remote programmingCOMMAND ERROR, PARAMETER ERROR, INVALID ITEM, INVALID VALUE PARAMETER, INVALID ENTER FORMAT, INVALID OUTPUT FORMAT, INVALID BLOCK FORMAT: IF STATEMENT, INVALID COMPARE OPERATOR, INVALID REPEAT MEM OVFL, INVALID REPEAT NEST LEVEL, INVALID ACTDEF, INVALID VARDEF, INVALID KEYNAME, INVALID KEYLBL, UNDF KEY, SOFTKEY OVFL
Traces, states and filesINVALID SAVE REG, INVALID REGISTER NUMBER, INVALID TRACE, INVALID TRACE NAME, INVALID TRACENAME, INVALID RANGE: Stop < Start, INVALID START INDEX, INVALID STOP INDEX, INVALID STORE DEST, INVALID FILENAME, INVALID <file name> NOT FOUND, INVALID CHECKSUM: USTATE, INVALID WINDOW TYPE, INTERNAL LOCKED, TABLE FULL, No points defined, Trace A is not available, Factory DLP, not adaptable
InterfaceINVALID HP-IB ADRS/OPERATION, INVALID HP-IB OPERATION REN TRUE, INVALID RS-232 ADRS/OPERATION
MeasurementsCannot reach N dB points, Too many signals with valid N dB points, Require 1/3/4 signals > PEAK EXCURSION above THRESHOLD, Signals do not fit expected % AM / TOI pattern, SIGNAL CLIPPED, SETUP ERROR, Marker Count Reduce SPAN, Marker Count Widen RES BW, Stop at marker not available with negative detection, INVALID DET: FM or TV option only, Required option not installed, Function not available in current Mode / with analog display
Time gatingCheck trigger input, Waiting for gate input, Verify gate trigger input is disconnected before CAL AMPTD

Hardware message on your 8590L? Analyzer Repair Inc. repairs and calibrates the HP / Agilent 8590L in-house in Cocoa, Florida. Tell us the message you see; estimates are free once the unit is in our lab. Contact us for service.

Message list based on the Agilent 8590 Series Assembly-Level Repair Service Guide (Chapter 14) and the 8590 E-Series and L-Series User's Guide.

HP / Agilent 8590L Replaceable Parts – click to expand

HP 8590L Replaceable Assemblies and Part Numbers

Assembly-level HP / Agilent replacement part numbers for the 8590L, taken from the 8590 Series assembly-level service guide. Parts marked with an option are only fitted to units that have that option. "Exchange" numbers are rebuilt units traded for your old assembly.

ReferenceAssemblyHP part numberOption
A1A1Keyboard08590-60201
A1A2Rotary Pulse Generator (old style)0960-0745
A1A2Rotary Pulse Generator (new style)1990-1525
A1R1Volume-Intensity Dual Pot Assembly08590-60041
A2Display08590-60377
A3Input Attenuator0955-0453
A4First Converter08590-60214
A5Second Converter08590-60117
A5A1Tripler08590-60155
A6YIG-Tuned Oscillator (YTO), 2-4 GHz0955-0974
A6A1YIG Board5062-4888
A7Analog Interface08590-60197
A7A1Tracking Generator Control08590-60196
A8Power Supply5062-8229
A8Power Supply Exchange08590-69005
A9Third Converter08591-60081
A9Third Converter08590-60344
A11Bandwidth Board08590-60398
A12Amplitude Control Board08590-60105
A13Bandwidth Board08590-60398
A14Log Amplifier Board08590-60386
A15Motherboard08590-60218
A16Processor Board08590-60416
A16A12X Memory Board08590-60399
A16A1BT1Battery1420-0394
A17Memory Card08590-60396
A2210 MHz Reference08591-60048Option 704
A24A1Tripler5086-7807
A24A2Tracking Converter08590-60146
A24A3Amplifier/Detector5086-7808
A24FL1Low-Pass Filter0955-0491
A25Counter Lock Board08591-60098
A25A1Sampler Assembly5086-7806
A40IB and Parallel Interface Board08590-60366Option 041
A40A1IB and Parallel Interface Connector08590-60368Option 041
A41RS-232 and Parallel Interface Board08590-60366Option 043
A41A1RS-232 and Parallel Interface Connector08590-60369Option 043
AT250 Ohm Termination0960-0053
B1Fan (includes cable assembly)5063-0243
F1Fuse 5.0 A 250 V F2110-0709
FL1Low-Pass Filter 4.9 GHz9135-0048
J1Input 50 Ω1250-2191
J1Input 75 Ω (MLA)0955-1005Option 001
J3RF OUT 50 Ω1250-2191Option 010
J3RF OUT 75 Ω (MLA)0955-1005Option 011

Part obsolete or unavailable? Many 8590L assemblies are no longer sold by Keysight. Analyzer Repair Inc. repairs these boards and modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

From the Agilent 8590 Series Analyzers Assembly-Level Repair Service Guide (08590-90316), Table 12-5.