Keysight (Agilent HP) E4407B Repair Service

E4407B ESA-E Spectrum Analyzer, 100 Hz to 26.5 GHz
Send us your Agilent HP E4407B Spectrum Analyzer for full repair and calibration. Our veteran-owned lab offers free estimates and quick turnaround.
•1.1 dB overall amplitude accuracyrn•100 Hz to 5 MHz RBW, 100 Hz to 300 Hz optionalrn•+7.5 dBm TOIrn•5 minute warm-up to guaranteed measurement accuracy
e4407b
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
E4407B Specifications – click to expand

E4407B Specifications

Frequency Span Resolution 2 Hz x N^4.

SpecificationValue
Frequency Range9 kHz to 26.5 GHz
Frequency Span Range0 Hz (zero span), 100 Hz to the range of the spectrum analyzer
Frequency Sweep Time1 ms to 4000 s, Span = 0 Hz 10 µs to 4000 s
Resolution Bandwidth (RBW)1 kHz to 5 MHz (–3 dB) in 1-3-10 sequence, 9 kHz and 120 kHz (-6 dB) EMI bandwidths
Amplitude Measurement RangeDisplayed average noise level to maximum safe input level
E4407B Options – click to expand

E4407B Options

1AX RS-232 and Parallel printer interface (includes RS-232 cable and BenchLink XL software) 1D5 High stability time base. 1D6 Time-gated spectrum analysis. 1D7 50 O/75 Omatching pad with dc block. 1DN 3 GHz tracking generator. 1DR Narrow resolution bandwidths. 1DS 3.0 GHz Preamplifier. A4J IF, sweep and video ports. AYX Fast zero-span sweeps. B72 Increases useable memory to 10 MB. BAA FM Demodulation/Deviation.

Agilent / HP E4407B Error Codes – click to expand

Agilent E4407B Error Codes and Messages

This E4407B error list covers the hardware, alignment and file messages from the ESA/PSA message manual, with notes on which E4407B assemblies to look at.

E4407B notes

  • The E4407B is an ESA-E analyzer to 26.5 GHz. This list uses the ESA numbering (108 is Cal Oscillator Unlock, 600-series file errors use ESA numbers).
  • Align Now, All on the E4407B needs the AMPTD REF OUT cabled to the INPUT (error 652 if it is missing).
  • Above 3 GHz the E4407B uses an A8A6 RYTHM (YIG-tuned harmonic mixer) with the A8A4 LOIS and A7A4 Frequency Extension board; preselector messages 208/209 apply there.
  • The E4407B accepts the E-series options such as Option 1D5 precision frequency reference, AYX fast ADC and BAA FM demod.

E4407B hardware and alignment errors

ErrorMessageMeaning
64Align All Now NeededThe stored alignment data is no longer valid. Run System, Alignments, Align Now, All. If the message comes straight back, or the alignment itself fails, there is a hardware fault in the signal path being aligned.
102Excessive 50 MHz RF input power is preventing internal alignmentsToo much 50 MHz energy is at the RF input for the internal alignment to run. Bring the input below 0 dBm and align again.
103Sampling Oscillator UnlockThe sampling oscillator that helps lock the first LO has lost lock. This is a hardware fault: on the E4407B the LO circuits are part of the A8 RF assembly (A8A1A2 RF Front End/LO bottom board with the YIG oscillator, and A8A1A1 Reference/Third Converter top board).
1041st LO UnlockThe first LO cannot hold phase lock. Usually hardware: on the E4407B the first LO is the YIG oscillator on the A8A1A2 RF Front End/LO board, locked to the A8A1A1 reference.
1052nd LO UnlockThe second LO has lost phase lock. On the E4407B check the A8A2 Second Converter and its 600 MHz drive (cable W10).
1064th LO UnlockThe fourth LO in the IF chain has lost phase lock. On the E4407B suspect the A3 IF assembly.
107Sample Clock UnlockThe ADC sample clock has lost lock. On the E4407B suspect the A3 IF assembly or its reference input.
108Cal Oscillator UnlockThe internal calibration oscillator has lost lock, so alignments and amplitude accuracy are affected. On the E4407B the calibrator is part of the A8 RF assembly (A8A1A2 RF Front End/LO bottom board with the YIG oscillator, and A8A1A1 Reference/Third Converter top board).
143Final IF OverloadThe final IF stage is overdriven. Add input attenuation or lower the signal. If it shows with no signal connected, suspect the IF section.
197Marker Freq Count HW FailureThe marker counter hardware did not return a valid count. A persistent failure points to the IF/counter circuitry.
206Unable to initialize flatness dataFlatness correction data in the internal EEROM could not be set up. Amplitude flatness will be wrong until the data is restored; this needs service.
207Unable to store flatness dataWriting flatness correction data to the internal EEROM failed. Usually means a memory problem or a failed adjustment; needs service.
622External reference missing or out of rangeExternal reference is selected but the signal at EXT REF IN is missing, too low, or not at the frequency entered. If it appears on internal reference, suspect the reference circuitry.
651Connect RF OUT to INPUTA tracking generator alignment was started without the TG RF OUT cabled to the analyzer INPUT. Connect it and repeat.
652Connect Amptd Ref Output to InputThe alignment needs the front-panel AMPTD REF OUT cabled to the INPUT on this model.
653Auto Align not available when using Calibration DefaultsCalibration defaults are loaded, so auto alignment is off until Align Now, All has been run successfully.
767Failed to Initialize ISTATE regions. Fatal LDS errorThe instrument state memory could not be initialized at power-up. Cycle power; if it repeats, the processor or its memory needs service.

E4407B status messages

Status messageMeaning
50 MHz Osc UnlevelThe internal 50 MHz amplitude reference is unleveled. Alignments cannot run correctly until this is fixed; points to the calibrator/reference circuitry.
First IF OverloadThe first IF is overdriven. Increase attenuation or reduce the input; with no input applied it indicates an RF/IF fault.
Frequency Reference Error (Freq Ref Unlocked)The frequency reference loop is unlocked or near unlock. On external reference, check the source matches the entered value; on internal reference it is a reference-board fault.
LO Unlock (Synth Unlocked)The LO phase-lock circuitry is unlocked. Readings are invalid until it is repaired.
LO Unlevel (LO Unleveled)The LO drive is unleveled inside the instrument, causing low or unstable signal levels.
ADC / IF / LO / RF Align FailureStatus bits that are set when a particular alignment routine fails. The failing routine identifies the section to troubleshoot.
Source LO Unlock / Source UnlevelTracking generator LO lost lock, or TG output cannot hold the set level. Check frequency and power are inside TG limits first; otherwise a TG board fault.

E4407B setup messages

ErrorMessageMeaning
129Meas UncalSweep is too fast for the RBW/VBW. Increase sweep time or press Auto Couple.
163Freq Count: Reduce Span/RBW ratioSpan is too wide for the RBW to count a marker. Narrow the span or widen the RBW.
201Option not installedThe function needs an option this unit does not have.
202No peak foundNo signal met the peak search criteria.
204TG Frequency LimitThe tracking generator has reached the end of its frequency range.
208Preselector is not available at this frequencyPreselector centering only works above about 3 GHz.
209Preselector centering failedCentering was attempted with too wide a span or no usable signal.
211RBW limited to 1 kHz when Span > 5 MHzDigital RBWs under 1 kHz are only available in narrower spans.
214TG start freq is less than 9 kHzTracking generator is uncalibrated below 9 kHz.
224Option not licensedThe option is present but has no license key installed.
751Instrument state set to initial valuesA saved or power-on state was corrupt and defaults were loaded.
757Unable to load user state, factory preset was doneThe user preset is invalid; save a good state as the user preset.
769Invalid instrument modeThe requested measurement personality is not installed.

E4407B file and storage errors

ErrorMessage
604File exists
605Media is protected
607File name error
609Media is not writable
612File not found
613Flash memory is full
614Bad, missing, or unformatted disk
615File is corrupt
617Wrong density floppy inserted

E4407B SCPI error ranges

RangeClass
−100 to −199Command errors (syntax, header)
−200 to −299Execution errors (setting out of range, not allowed now)
−300 to −399Device-specific errors (queue overflow, self-test, memory)
−400 to −499Query errors (unterminated or interrupted query)

LO unlock, alignment failure or IF overload on your E4407B? Analyzer Repair Inc. repairs and calibrates the Agilent E4407B in-house in Cocoa, Florida, including YTO, RF assembly, synthesizer and IF board faults. Free estimate once the unit is in our lab. Contact us for service.

Message numbers from Agilent ESA/PSA Instrument Messages and Functional Tests (E4440-90619); assembly references from the Agilent ESA Spectrum Analyzers Service Guide.

Agilent / HP E4407B Replaceable Parts – click to expand

Agilent E4407B Replaceable Assemblies and Part Numbers

Board and RF module part numbers for the Agilent E4407B. Exchange and refurbished numbers are factory-rebuilt units traded for the old assembly.

ReferenceAssemblyAgilent part numberNotes
A1A1Front panel interface board kitE4401-60241
A1A1A1Rotary pulse generator (RPG)1990-1865
A1A2LCD assembly kitE4401-60242Serial prefix US4119 and above (earlier units take E4401-60241 and E4401-60242)
A1A2DS1/DS2Display backlight2090-0577Upper and lower
A1A3Front frame assemblyE4401-60257Except Option 060
A1MP5Main keypadE4401-40002
A2A13.0 GHz TG driver board with shieldsE4403-60043Option 1DN (exchange E4403-69043)
A2A23.0 GHz BITG microcircuit5086-7993Option 1DN (exchange 5086-6993)
A2A33.0 GHz TG output attenuator0955-0679Option 1DN
A3IF assembly with shieldsE4401-60183
A4Processor assembly, enhanced memoryE4401-60233Option B72
A4A1Flash SIMM, 4 MBE4401-6010512 MB E4401-60081 with Option B72
A4A2DRAM SIMM, 32 MBE4401-60289Option B72
A4MAU58Lithium battery, 3 V1420-0556
A5Power supply assemblyE4401-60186
A5B1Fan3160-0866
A6Speaker/floppy assemblyE4401-60008
A7Motherboard, ESA-E seriesE4401-60107
A7A1IF/video/sweep portsE4401-60096Option A4J
A7A2GPIB assemblyE4401-60013Option A4H
A7A3RS-232 assemblyE4401-60014Option 1AX
A7A4Frequency extension assemblyE4404-60001
A7A5Fast ADC assemblyE4401-60083Option AYX
A7A6FM demod assemblyE4401-60343Option BAA
A7A7Digital demod DSP assemblyE4401-60088Option B7D
A7A8Digital demod RF assemblyE4401-60285Option B7E
A7A9ACPR dynamic range extensionE4403-60065Option 120
A7A10Bluetooth FM demodulatorE4403-60223
A7A12IF out assemblyE4403-60076Option H55
A7A12IF out assemblyE4404-60034Option H70
A83 GHz RF complete assemblyE4403-60073Without precision reference (exchange E4403-69073)
A83 GHz RF complete assemblyE4403-60074With Option 1D5 precision reference (exchange E4403-69074)
A8A1A1Reference/third converter (top board)E4403-60001Without precision reference
A8A1A1Reference/third converter (top board)E4403-60053Option 1D5
A8A1A23 GHz RF front end/LO (bottom board)E4403-60033
A8A1A2MC1YIG oscillator0955-1039
A8A2Second converter5086-7958
A8A3DC blockE4403-60026Except Option UKB
A8A4LO amplifier/IF switch (LOIS)E4404-60030Exchange E4404-69030
A8A5Input attenuator, 26.5 GHz33321-60057Except Option UKB (UKB: 33321-60061)
A8A6YIG-tuned harmonic mixer (RYTHM), 26.5 GHz5087-7031Exchange 5087-6031
A8FL1Low pass filter, 3.1 GHz0955-0988
A8J1RF input Type-N connector kitE4403-60015Except Option 060
A8J1RF input APC-3.5 connector kitE4407-60002Option BAB

Part obsolete? Many E4407B assemblies are no longer sold new. Analyzer Repair Inc. repairs E4407B RF assemblies, YTOs, IF and synthesizer boards at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

From the Agilent ESA Spectrum Analyzers Service Guide.

Located in: Spectrum Analyzers