Keysight (Agilent HP) E5070A Repair Service
Product model
E5070A Network Analyzer 300kHz-3GHzServices
Description

Repair prices
E5070A Specifications – click to expand
E5070A Specifications
- ENA-series vector network analyzer
- Windows-based with color display, LAN, USB and GPIB interfaces
| Specification | Value |
|---|---|
| Frequency range | 300 kHz to 3 GHz |
E5070A Options – click to expand
E5070A Options
- 213/214, 2-Port. 313/314, 3-Port. 413/414, 4-Port (xx4 adds bias tees)
- 008, Frequency Offset
- 1E5, High-Stability Frequency Reference
Agilent / HP E5070A Error Codes – click to expand
Agilent E5070A Error Messages
The E5070A messages below are the ones that point to hardware, followed by what a failed performance test means on this model. Errors show on a red bar at the lower left; numbered ones can be read over GPIB.
E5070A notes
- The E5070A runs a power-on self test at every start; a failure is reported as error 241 and broken down by the internal test in the service menu.
- If the rear fan does not run, the A50 power supply is suspect; check the +3.3 V / +5 V monitor LEDs on the A21 board.
E5070A hardware errors
| Error | Message | Meaning |
|---|---|---|
| 220 | Phase lock loop unlocked | A synthesizer PLL lost lock during measurement. The service manual points to the A1 source board (small chance: the ADC section of the A2 receiver). |
| 241 | Power on test failed | The power-on self test found a hardware fault. Run the internal test from the service menu; the first failed test identifies the board (A1 source, A2 receiver (A14 RF / A17 IF boards), A3 level vernier, A4 power amp, A6 switch/temperature control, A7 attenuator). |
| 240 | Calibration data lost | The system calibration file is missing or damaged at startup. The analyzer needs its adjustment data restored or regenerated. |
| 221 | Port 1 receiver overload | Too much power into port 1. If it appears with nothing connected, the analyzer is faulty. |
| 222 | Port 2 receiver overload | Same as 221 for port 2. |
| 223 | Port 3 receiver overload | 3- and 4-port options only. |
| 224 | Port 4 receiver overload | 4-port options only. |
| — | Fatal Error: Failed to Initialize DSP / DSP Driver | The A21 analog interface board or A20 digital motherboard did not start. Also shown after a failed firmware update. |
| — | Fatal Error: Failed to Update DSP Code | Writing DSP firmware to flash on the A21 analog interface board failed; reinstall firmware, then suspect A21 analog interface board or A20. |
E5070A calibration, file and setup errors
| Error | Message |
|---|---|
| 20 | Additional standard needed |
| 22 | Calibration method not selected |
| 30 | Valid ECal module not found |
| 31 | Failed to configure ECal module |
| 41 | Peak not found |
| 53 | Log sweep requires 2 octave minimum span |
| 100 | Failed to read file |
| 101 | Failed to write file |
| 105 | Recall failed |
| 106 | Save failed |
| 120 | Printer error |
| 200 | Option not installed |
SCPI error ranges
| Range | Class |
|---|---|
| −100 to −199 | Command errors |
| −200 to −299 | Execution errors |
| −300 to −399 | Device-specific errors |
| −400 to −499 | Query errors |
Hardware error on your E5070A? Analyzer Repair Inc. repairs and calibrates the Agilent E5070A in-house in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.
Based on the Agilent E5070A/E5071A ENA Service Manual (E5070-90110).
Agilent / HP E5070A Replaceable Parts – click to expand
Agilent E5070A Replaceable Assemblies and Part Numbers
Assembly-level part numbers for the E5070A. Check your options before ordering option-only parts.
| Reference | Assembly | Agilent part number | Notes |
|---|---|---|---|
| A1 | Source board | E5071-66551 | Exchange E5071-69551 |
| A7 | Source step attenuator | 33321-60059 | Options 214, 314, 414 |
| A8 | RF switch, SPDT | 5087-7137 | Options 31x, 41x |
| A14 | Receiver RF board | E5071-61014 | Exchange E5071-69014 |
| A17 | Receiver IF board | E5071-66517 | |
| A5 | Crystal oven (high stability reference) | E4991-66505 | Option 1E5 |
| A6 | Switch control board | E5070-66586 | |
| A20 | Digital motherboard | E5070-61030 | |
| A21 | Interface board | E5070-66521 | |
| A22 | Front panel keyboard | E5070-66522 | |
| A23 | Handler I/O board | E5070-66523 | |
| A26 | LCD interface card | 54810-66525 | |
| A27 | Hard disk drive | E5070-65030 | |
| A28 | Floppy disk drive | 0950-3334 | |
| A50 | ATX power supply | 0950-3961 | |
| A52 | Inverter | 0950-4091 | |
| — | Analog motherboard | E5070-66510 | |
| — | LCD assembly | E5070-60101 | Touch-panel LCD E5070-60102 with Option 016 |
Part obsolete? Many E5070A assemblies are no longer sold new. Analyzer Repair Inc. repairs E5070A boards and RF modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.
From the Agilent E5070A/E5071A ENA Service Manual (E5070-90110).