Keysight (Agilent HP) 8590A Repair Service
Product model
8590A Portable RF Spectrum Analyzer, 10 kHz - 1.5 GHzServices
Description

Repair prices
8590A Specifications – click to expand
8590A Specifications
With a wide frequency range (10kHz to 1.5GHz, 4-digit resolution), large amplitude range (-115 dBm to +30 dBm), and a 50-ohm input,
8590A Options – click to expand
8590A Options
001 - 75 Ohm Input Impedance
HP 8590A Error Messages – click to expand
HP 8590A Error Messages
The HP 8590A portable spectrum analyzer runs self-tests at power-up and every time PRESET is pressed. When a test fails, a message appears on screen. These are the hardware messages from the 8590A service manual and the boards usually at fault.
First step: connect CAL OUTPUT to the RF input, run CAL FREQ and CAL AMPTD, and press CAL STORE. Many 8590A messages come from bad calibration data.
8590A hardware messages
| Message | Meaning | Suspect assemblies |
|---|---|---|
| ADC-GND FAIL | Processor analog-to-digital converter failure. | A16 processor |
| ADC-TIME FAIL | Processor analog-to-digital converter failure. | A16 processor |
| ADC-2V FAIL | Processor analog-to-digital converter failure. | A16 processor |
| CAL: FM SPAN SENS FAIL | Could not set the FM coil span sensitivity. | A7 analog interface (FM coil driver) |
| CAL: SPAN SENS FAIL | The YIG main coil sensitivity calibration failed. | A7 analog interface (main coil driver) |
| CAL: LINEAR DET FAIL | The linear detector calibration failed. | Attenuator cal constant missing or overwritten, or A12 step gain / A14 log amp |
| CAL: RES BW AMPL FAIL | Relative loss of a resolution bandwidth filter is wrong. | RF/IF section |
| FAIL: xxxx xxxxxxxxxx | Power-up self-test found a circuit fault (see below). | A16 processor and bus |
8590A self-calibration progress (M)
While CAL routines run, the 8590A shows progress messages such as CAL: MC DELAY, FM DELAY, SWEEP, SWP DELAY, FREQ, SPAN, AMPTD, 3dB BW, ATTEN, LOG AMP and DONE. These are normal.
8590A setup and information messages
| Message | Type | Meaning |
|---|---|---|
| MEAS UNCAL | U | Sweep time, span and bandwidth settings make the measurement uncalibrated. Press AUTO COUPLE or adjust the settings. |
| SRQ 110 | M | Hardware broken service request. Look for the hardware message shown with it. |
| SRQ 140 | M | An illegal command was received over the bus. |
| INVALID SAVE REG | U | Nothing has been saved in that state register. |
| INVALID SYMTAB: EEROM OVERFLOW | U | Too much data has been cataloged into memory. |
| INVALID SYMTAB: SYMTAB OVERFLOW | U | The symbol table is full; too many user-defined items in memory. |
8590A remote programming messages (U)
These come from bad remote commands over HP-IB, HP-IL or RS-232, not from hardware: COMMAND ERROR, PARAMETER ERROR, CONFLICT TABLE OVERFLOW, INVALID AUNITS, INVALID DETECTOR, INVALID TRIGGER MODE, INVALID TRACE, INVALID TRACE NAME, INVALID VARDEF, INVALID KEYNAME, INVALID WINDOW TYPE, INVALID ENTER FORMAT, INVALID OUTPUT FORMAT, INVALID BLOCK FORMAT: IF STATEMENT, INVALID COMPARE OPERATOR, INVALID CHECKSUM: USTATE, INVALID REPEAT MEM OVFL, INVALID REPEAT NEST LEVEL, INVALID HP-IB ADDRESS OR OPERATION, INVALID HP-IB OPERATION REN TRUE, INVALID HP-IL ADDRESS OR OPERATION, INVALID HP-IL OPERATION REN TRUE, INVALID RS-232 ADDRESS OR OPERATION, SOFTKEY OVFL, UNDEF KEY.
8590A power-up FAIL codes
The 4-digit part of a FAIL code shows which circuit on the A16 processor board failed (ROM, RAM, EEPROM, I/O bus, CPU or video RAM), and matches the row of test LEDs on A16. The longer second part breaks down into the I/O address bus, the I/O data bus to the A7 analog interface, and the video board RAM.
FAIL: 1330 0000000000 is a common 8590A code. It usually means bad data in the EEPROM that holds calibration constants, with no bus or video RAM fault. Rerunning CAL FREQ and CAL AMPTD and pressing CAL STORE usually restores normal operation. If it keeps coming back, the EEPROM or its data needs service.
8590A calibration data problems
Lost or altered calibration constants can cause spurious signals, LIN DET FAIL, SRQ 110, or codes such as 1330 0000000000 and 5300 0000000000. If the A3 attenuator has been replaced, new calibration constants must be generated and entered.
Need the 8590A fixed? We repair older 8590 series analyzers to the board and component level and return them calibrated. Free estimate once it arrives; you approve before any work. Contact us for service.
Based on the HP 8590A Installation Manual (Appendix A) and the HP 8590A Support Manual (Chapter 4).