Keysight (Agilent HP) 8753ES Repair Service

8753ES network analyzer 30kHz-3GHz
Keep your Agilent HP 8753ES Network Analyzer in service with our repair and calibration. We specialize in unsupported equipment. Free evaluations.
Keysight Agilent HP 8753ES vector network analyzers perform complete characterization of RF components.
8753es
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
8753ES Specifications – click to expand

8753ES Specifications

  • 8753ES Option 011 Receiver
  • Description Specification Supplemental Information
  • Frequency Range
  • Option 011 300 kHz to 3.0 GHz 300 kHz to 3 GHz, typ
  • Option 011 and 006 30 kHz to 6.0 GHz 10 kHz to 6 GHz, typ
  • Frequency Response
  • 300 kHz to 3 GHz ±1 dB at −10 dBm
  • 3 GHz to 6 GHz ±2 dB at −10 dBm
  • Frequency Response (Ratio)
  • Amplitude
  • 300 kHz to 3 GHz ±0.5 dB, typ
  • 3 GHz to 6 GHz ±2.0 dB, typ
  • Phase
  • 300 kHz to 3 GHz ±3° −10 dBm, all inputsa
  • 3 GHz to 6 GHz ±10° −10 dBm, all inputsa
  • Stability (Ratio Measurement)
  • Magnitude
  • 300 kHz to 3 GHz 0.02 dB/°C, typ
  • 3 GHz to 6 GHz 0.04 dB/°C, typ
  • Phase
  • 300 kHz to 3 GHz 0.05 deg/°C, typ
  • 3 GHz to 6 GHz 0.20 deg/°C, typ
  • Impedance
  • Option 011 50 Ω, nominal
  • 30 kHz to 300 kHz 7 dB, typ
  • 300 kHz to 2 MHz 20 dB
  • 2 MHz to 1.3 GHz 24 dB
  • 1.3 GHz to 3 GHz 19 dB
  • 3 GHz to 6 GHz 15 dB
8753ES Options – click to expand

8753ES Options

  • 006, 6 GHz Frequency Extension
  • 010, Time Domain
  • 011, Receiver Configuration
  • 014, Configurable Test Set
  • 075, 75 Ohm System Impedance
  • 1D5, High-Stability Frequency Reference
HP / Agilent 8753ES Error Messages – click to expand

HP / Agilent 8753ES Error Messages

When the HP / Agilent 8753ES S-parameter network analyzer (30 kHz to 3 GHz (6 GHz with Option 006)) detects a problem, it shows a message and an error number on screen. These are the service-related messages from the 8753ES service guide, explained in plain language.

8753ES notes

  • The 8753ES has a built-in S-parameter test set (transfer switch, two test port couplers and R, A and B samplers).
  • Option 006 extends the 8753ES to 6 GHz; Option 075 changes it to 75 ohm test ports.
  • Phase lock errors on the 8753ES usually point to the source (YIG oscillator or source assembly) or the phase lock board, but check R channel input level first.

8753ES hardware error messages

MessageError no.TypeMeaning
NO IF FOUND: CHECK R INPUT LEVEL5HThe first IF was not detected while pretuning the phase lock. Usually a source fault, or the R channel is getting too little signal (needs at least −35 dBm).
NO PHASE LOCK: CHECK R INPUT LEVEL7HThe IF was found during pretune but the loop never locked. Check R channel level, then the source and phase lock circuits.
PHASE LOCK LOST8HPhase lock was acquired and then dropped during the sweep.
PHASE LOCK CAL FAILED4HThe automatic phase lock calibration (run at power-on or when pretune drifts) failed.
POSSIBLE FALSE LOCK6HThe source may be locked to the wrong harmonic comb tooth, so frequencies can be wrong.
SWEEP TIME TOO FAST12HThe fractional-N and digital IF circuits lost synchronization; points to the digital control section.
POWER UNLEVELED179HSource output is not leveled. Can be a power setting beyond the source range or a source/ALC fault.
POWER SUPPLY HOT!21HThe power supply is overheating. Check the fan, filter and air flow.
POWER SUPPLY SHUT DOWN!22HOne or more supplies shut down, usually from an overload or short on a board.
PROBE POWER SHUT DOWN!23HThe front-panel probe power supply shut down from a short or overload.
BATTERY FAILED. STATE MEMORY CLEARED183HThe memory backup battery has failed and stored instrument states were cleared. Replace the battery on the CPU board.
BATTERY LOW! STORE SAVE REGS TO DISK184HThe memory backup battery is getting weak. Save your states to disk and replace the battery.
CORRECTION CONSTANTS NOT STORED3HWriting correction constants to EEPROM failed. The A9 CPU write-protect switch/jumper must be set to allow writing.
DEADLOCK111HAn internal firmware deadlock occurred. Cycle power; if it repeats, the CPU or a firmware problem is likely.
DISK HARDWARE PROBLEM39HThe disk drive is not responding properly.
DISK READ/WRITE ERROR189HA disk read or write failed; try another disk, then suspect the drive.
DISK MESSAGE LENGTH ERROR190HThe disk drive and analyzer are out of step during a transfer.
INITIALIZATION FAILED47HThe disk could not be initialized; check the disk and drive.

8753ES calibration, setup and information messages

MessageError no.TypeMeaning
OVERLOAD ON INPUT R, POWER REDUCED57UToo much power at the R input; source power was reduced (P⇓).
SOURCE POWER TURNED OFF, RESET UNDER POWER MENUMAn input overload tripped the source off. Remove the overload and turn power back on in the power menu.
CALIBRATION REQUIRED63UNo calibration set matches the current stimulus or measurement; perform a new cal.
CURRENT PARAMETER NOT IN CAL SET64UThe parameter you selected was not included in the active calibration.
CORRECTION TURNED OFF66USweep type, start, span or number of points changed from the cal, so correction was switched off.
NO CALIBRATION CURRENTLY IN PROGRESS69UA cal step was requested but no calibration is running.
NO SPACE FOR NEW CAL. CLEAR REGISTERS70UMemory is full; clear saved registers to store a new calibration.
CALIBRATION ABORTED74UThe active channel was changed during a cal, so it stopped. Restart on the right channel.
WRONG DISK FORMAT, INITIALIZE DISK77UThe disk is not in LIF format; initialize it before use.
TROUBLE! CHECK SETUP AND START OVER115UThe test equipment setup for an adjustment is not correct; check the setup diagram.
TEST ABORTED113UA service test or adjustment was stopped before it finished.
POWER METER INVALID116UThe power meter reading is out of range or not valid.
PWR MTR: NOT ON/CONNECTED OR WRONG ADDRS117UThe power meter is off, not connected by GPIB, or set to the wrong address.
POWER METER NOT SETTLED118UPower meter readings are not stable.
NOT ALLOWED DURING POWER METER CAL198UThat function cannot be used while a source power meter calibration is running.
INSUFFICIENT MEMORY, PWR MTR CAL OFF154UNot enough memory for power meter cal data, so it was turned off.
SAVE FAILED. INSUFFICIENT MEMORY151UNot enough memory to save the state; clear some registers.
NOT ENOUGH SPACE ON DISK FOR STORE44UThe disk is full.
NO FILE(S) FOUND ON DISK45UNo files of the requested type are on the disk.
PARALLEL PORT NOT AVAILABLE FOR COPY167UThe parallel port is set for GPIO, so it cannot be used for printing.
PARALLEL PORT NOT AVAILABLE FOR GPIO165UThe parallel port is set for printing, so it cannot be used as GPIO.
SWEEP MODE CHANGED TO CW TIME SWEEP187MThe sweep mode was switched to CW time sweep.

Hardware error on your 8753ES? Analyzer Repair Inc. repairs and calibrates the HP / Agilent 8753ES in-house in Cocoa, Florida, including source, phase lock, sampler and test set faults. Free estimate once the unit is in our lab. Contact us for service.

Message list based on the Agilent 8753ES/ET Service Guide (08753-90484), service error messages. Measurement-only messages are in the 8753ES reference guide.

HP / Agilent 8753ES Replaceable Parts – click to expand

HP 8753ES Replaceable Assemblies and Part Numbers

Assembly-level HP / Agilent part numbers for the 8753ES network analyzer, from the Agilent 8753ES/ET Service Guide (08753-90484). Option-specific parts are marked in the Notes column, with rebuilt-exchange numbers where HP offered them.

ReferenceAssemblyHP part numberNotes
A1Front panel board08720-60267
A2Front panel interface board08753-60358
A3Source assembly, 3 GHz08753-60231Standard (exchange 08753-69231)
A3Source assembly, 6 GHz08753-60146Option 006 (exchange 08753-69146)
A4Sampler R08753-60907
A5Sampler A08753-60908
A6Sampler B08753-60908
A7Pulse generator board08753-60164
A8Post regulator board08753-60949
A9CPU repair kit08753-60315Exchange 08753-69315
A9BT1Battery, lithium 3 V1420-0338
A10Digital IF board08753-60958
A11Phase lock board08753-60162
A12Reference board08753-60957
A13Frac-N analog board08753-60013
A14Frac-N digital board08753-60068
A15Preregulator0950-3488
A16Rear panel interface board08720-60138
A17Motherboard08753-60360
A18LCD replacement assembly08753-60937
A19Graphics processor (GSP) board08753-60359
A20Disk drive0950-2782
A21, A22Test port coupler5087-7007Exchange 5087-6007
A21, A22Test port coupler, 75 ohm5087-7008Option 075 (exchange 5087-6008)
A23LED front panel board08753-60145
A24Transfer switch, solid state5087-7067Exchange 5087-6067
A25Test set interface board08753-60280
A26High stability frequency reference08753-60158Option 1D5
A27Inverter0950-3379
B1Fan08415-60036
RPGRPG knob assembly with cable1990-1864

Part obsolete? Many 8753ES assemblies are no longer sold new. Analyzer Repair Inc. repairs 8753ES boards, samplers and RF modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

From the Agilent 8753ES/ET Service Guide (08753-90484).