Keysight (Agilent HP) E4447A Repair Service

E4447A PSA Spectrum Analyzer, 3 Hz - 42.98 GHz
Expert repair and calibration for your Agilent E4447A Spectrum Analyzer. We specialize in obsolete, unsupported equipment. Free evaluations.
The Agilent E4447A PSA high-performance spectrum analyzer measures and monitors complex RF, microwave, and millimeter-wave signals up to 42.98 GHz. Performancern+/-0.19 dB absolute amplitude accuracyrn-154 dBm displayed average noise level (DANL)rn-118 dBc/Hz phase noise at 10 kHz offset 81 dB W-CDMA ACLR dynamic range Analysis bandwidth 10 MHz analysis bandwidth (Option B7J for the Basic mode)
e4447a
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
E4447A Specifications – click to expand

E4447A Specifications

SpecificationValue
Frequency range3 Hz to 42.98 GHz (DC coupled)
Sweep TimeSpan >= 10 Hz, 1 ms to 2000 s
Phase Noise at 1GHz (10 kHz Offset)-116 dBc/Hz
Maximum Third Order Dynamic Range, 1 GHz113 dB
Displayed Average Noise at 1 GHz-168 dBm
Standard Attenuator Range/Step70 dB / 2 dB
Overall Amplitude Accuracy± 0.62 dB
Resolution Bandwidth1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz
Standard Analysis Bandwidth10 MHz
E4447A Options – click to expand

E4447A Options

  • Opt 226, Phase noise
  • Opt 219, Noise figure Requires Option IDS or 110 to meet specifications
  • Opt 241, Flexible digital modulation analysis
  • Opt BAF, W-CDMA Requires B7J
  • Opt 210, HSDPA/HSUPA (for W-CDMA) Requires B7J and BAF
  • Opt 202, GSM w/ EDGE Requires B7J
  • Opt B78, cdma2000 Requires B7J
  • Opt 214, 1xEV-DV Requires B7J and B78
  • Opt 204, 1xEV-DO Requires B7J
  • Opt BAC, cdmaOne Requires B7J
  • Opt BAE, NADC, PCD Requires B7J
  • Opt 217, WLAN Requires 122 or 140
  • Opt 211, TD-SCDMA power measurement
  • Opt 212, TD-SCDMA modulation
  • Opt 213, HSPA for TD-SCDMA Requires Option 212
  • Opt 215, External source control
  • Opt 266, Programming code compatibility suite
  • Opt 233, Built-in measuring receiver personality
  • Opt 23A, AM/FM/PM triggering Requires Option 233
  • Opt 23B, CCITT filter Requires Option 233
  • Opt 239, N9039A RF preselector control. Options
  • Opt 1DS, RF internal preamplifier (100 kHz to 3 GHz), Excludes 110
  • Opt 110, RF/µW internal preamplifier (10 MHz to upper frequency limit of the PSA), Excludes 1DS
  • Opt B7J, Digital demodulation hardware
  • Opt 123, Switchable MW preselector bypass Excludes AYZ
  • Opt 124, Y-axis video output
  • Opt AYZ, External mixing, excludes 123
  • Opt 107, Audio input 100 k-ohm Requires 233 to operate; excludes 122, 140
  • Opt 111, USB device side I/O interface Shipped standard in all PSA instruments with serial number prefix >= MY4615 unless 117 license is activated
  • Opt 115, 512 MB user memory
  • Opt 117, Secure memory erase Excludes 115
  • Opt BAB, Replaces type-N input connector with APC 3.5 connector
  • Opt HYX, 21.4 MHz IF output Available for all PSA models
Agilent / HP E4447A Error Codes – click to expand

Agilent E4447A Error Codes and Messages

Below are the E4447A error numbers that matter most for repair, followed by the status annunciators and the setup and file messages. Meanings are written for this PSA model.

E4447A notes

  • The E4447A is a PSA-series analyzer to 42.98 GHz. PSA and ESA share one message manual, but some numbers mean different things (for example 108, and the 600-series file errors), so this list uses the PSA numbering.
  • The E4447A millimeter front end uses the A19 SBTX/RYTHM assembly, the A29 SBTX Driver board and the A30 FIFA first IF amplifier, with a 10 dB/switch (A14) and 60 dB (A15) input attenuator pair.
  • With Option 123 fitted, error 235 means the 26.5–50 GHz preselector bypass is on, not a fault.
  • The E4447A uses its own A10 3rd Converter (E4447-60002); fitting the standard converter gives the "Incompatible Hardware found for E4447A: Third Converter" message and a 26.5 GHz limit.
  • Errors 787–789 on the E4447A only apply with the Option 122/140 wideband IF installed.

E4447A hardware and alignment errors

ErrorMessageMeaning
64Align All Now NeededThe stored alignment data is no longer valid. Run System, Alignments, Align Now, All. If the message comes straight back, or the alignment itself fails, there is a hardware fault in the signal path being aligned.
102Excessive 50 MHz RF input power is preventing internal alignmentsToo much 50 MHz energy is at the RF input for the internal alignment to run. Bring the input below 0 dBm and align again.
103Sampling Oscillator UnlockThe sampling oscillator that helps lock the first LO has lost lock. This is a hardware fault: on the E4447A check the A12A2 Sampling Oscillator board and the 600 MHz reference reaching it from A11 over cable A12W2.
1041st LO UnlockThe first LO cannot hold phase lock. Usually hardware: on the E4447A the usual suspects are the A18 YTO, the A12A1 LO/Synthesizer board, the A21 FELOMA (LO multiplier/amplifier) and the A13 Front End Driver (YTO coil drive and LO leveling).
1052nd LO UnlockThe second LO has lost phase lock. On the E4447A the 2nd LO is on the A9 2nd LO/Fan Control assembly, fed with 600 MHz from the A11 Reference.
1064th LO UnlockThe fourth LO in the IF chain has lost phase lock. On the E4447A look at the IF chain (A8 Analog IF / A7 Digital IF) and its reference inputs.
107Sample Clock UnlockThe ADC sample clock has lost lock. On the E4447A the sample clock serves the A7 Digital IF; check it and the reference feeding it.
143Final IF OverloadThe final IF stage is overdriven. Add input attenuation or lower the signal. If it shows with no signal connected, suspect the IF section.
197Marker Freq Count HW FailureThe marker counter hardware did not return a valid count. A persistent failure points to the IF/counter circuitry.
206Unable to initialize flatness dataFlatness correction data in the internal EEROM could not be set up. Amplitude flatness will be wrong until the data is restored; this needs service.
207Unable to store flatness dataWriting flatness correction data to the internal EEROM failed. Usually means a memory problem or a failed adjustment; needs service.
622External reference missing or out of rangeExternal reference is selected but the signal at EXT REF IN is missing, too low, or not at the frequency entered. If it appears on internal reference, suspect the reference circuitry.
653Auto Align not available when using Calibration DefaultsCalibration defaults are loaded, so auto alignment is off until Align Now, All has been run successfully.
767Failed to Initialize ISTATE regions. Fatal LDS errorThe instrument state memory could not be initialized at power-up. Cycle power; if it repeats, the processor or its memory needs service.
605Data corrupt or stale; EEPROM copy of A calibration file in EEPROM is corrupt beyond repair by its error-correction codes, so default values were loaded. Amplitude and frequency accuracy may be off; frequent occurrences point to failing memory.
606Data questionable; EEPROM copy of The EEPROM copy of an internal file is unusable and a default was used in its place. Report for service.
608Mass storage error; EEPROM write timeoutAn EEPROM write did not finish in time. Repeated occurrences indicate a memory or board fault.
609Restoration of NVRAM dataA file in non-volatile RAM was repaired from its error-correction codes. Harmless once; if it appears at every power-up, the memory is failing.
786Cal file not initialized correctly, An older LO board is fitted whose EEPROM band-edge cal file has no millimeter-band values. Seen in service mode after board swaps.
787WBIF Self Test FailedThe Option 122/140 wideband IF failed its power-on loop-back test (ADC image cal, DAC playback, power and S/N check). Hardware fault or incorrect installation of the wideband IF boards.
788WBIF ADC Image Calibration FailedThe two wideband ADCs could not be resynchronized. This calibration only fails with a hardware problem in the Option 122/140 digitizer.
789WBIF Firmware ResetThe local processor on the wideband digitizer was reset. Repeated resets indicate a wideband IF hardware fault.

E4447A status messages

Status messageMeaning
50 MHz Osc UnlevelThe internal 50 MHz amplitude reference is unleveled. Alignments cannot run correctly until this is fixed; points to the calibrator/reference circuitry.
First IF OverloadThe first IF is overdriven. Increase attenuation or reduce the input; with no input applied it indicates an RF/IF fault.
Frequency Reference Error (Freq Ref Unlocked)The frequency reference loop is unlocked or near unlock. On external reference, check the source matches the entered value; on internal reference it is a reference-board fault.
LO Unlock (Synth Unlocked)The LO phase-lock circuitry is unlocked. Readings are invalid until it is repaired.
LO Unlevel (LO Unleveled)The LO drive is unleveled inside the instrument, causing low or unstable signal levels.
ADC / IF / LO / RF Align FailureStatus bits that are set when a particular alignment routine fails. The failing routine identifies the section to troubleshoot.
Incompatible Hardware found for E4447A: Third ConverterAt power-up the E4447A found a third converter that is not the E4447A part. It runs but is limited to 26.5 GHz until the correct converter is fitted.

E4447A setup messages

ErrorMessageMeaning
129Meas UncalSweep is too fast for the RBW/VBW. Increase sweep time or press Auto Couple.
163Freq Count: Reduce Span/RBW ratioSpan is too wide for the RBW to count a marker. Narrow the span or widen the RBW.
201Option not installedThe function needs an option this unit does not have.
202No peak foundNo signal met the peak search criteria.
208Preselector is not available at this frequencyPreselector centering only works above about 3 GHz.
209Preselector centering failedCentering was attempted with too wide a span or no usable signal.
211RBW limited to 1 kHz when Span > 5 MHzDigital RBWs under 1 kHz are only available in narrower spans.
224Option not licensedThe option is present but has no license key installed.
2293 – 26.5 GHz preselector bypassed, can't adjustOption 123 preselector bypass is on, so the microwave preselector cannot be centered.
23526.5 – 50 GHz preselector bypassed, can't adjustOption 123 bypass is on for the millimeter band, so it cannot be centered.
236License conflictTwo options that cannot be licensed together were requested.
751Instrument state set to initial valuesA saved or power-on state was corrupt and defaults were loaded.
757Unable to load user state, factory preset was doneThe user preset is invalid; save a good state as the user preset.

E4447A file and storage errors

ErrorMessage
601File not found
602Media is protected
603File is corrupt
604File name error
613Flash memory is full
615File exists
617Wrong density floppy inserted
627Media is not writable
763Bad, missing, or unformatted disk

E4447A SCPI error ranges

RangeClass
−100 to −199Command errors (syntax, header)
−200 to −299Execution errors (setting out of range, not allowed now)
−300 to −399Device-specific errors (queue overflow, self-test, memory)
−400 to −499Query errors (unterminated or interrupted query)

LO unlock, alignment failure or IF overload on your E4447A? Analyzer Repair Inc. repairs and calibrates the Agilent E4447A in-house in Cocoa, Florida, including YTO, RF assembly, synthesizer and IF board faults. Free estimate once the unit is in our lab. Contact us for service.

Message numbers from Agilent ESA/PSA Instrument Messages and Functional Tests (E4440-90619); assembly references from the Agilent PSA Service Guide (E4440-90616).

Agilent / HP E4447A Replaceable Parts – click to expand

Agilent E4447A Replaceable Assemblies and Part Numbers

Assembly-level part numbers for the E4447A PSA spectrum analyzer, taken from the Agilent PSA Service Guide (E4440-90616). Confirm your options and serial prefix before ordering.

ReferenceAssemblyAgilent part numberNotes
A1Flat panel display2090-0897Serial prefix 4611 and above
A1Flat panel display retrofit kitE4440-60445Below serial prefix 4611 (replaces 2090-0396)
A1A1Display converter boardE4440-63265
A2Front panel interface (with inverter boards and cable)E4440-63266
A2A1/A2A2Inverter boards0950-3379
A3Keyboard assemblyE4406-60004Keypad not included
A5Power supply assembly0950-4447
A6SCSI interface boardE4406-60065Replaced by A39 on serial prefix 4611 and above
A7Digital IF assemblyE4440-60206Serial prefix 4611 and above; also Option 122/140, 124
A7Digital IF retrofit kitE4440-60711Replaces E4440-60025 / E4440-60195
A8Analog IF assemblyE4440-60256
A92nd LO / fan controlE4440-60431
A103rd converter (with W18)E4447-60002
A11Reference assemblyE4440-60225
A12Synthesizer assembly (A12A1, A12A2, shields, cables)E4440-60185Firmware A.04.08 or later required
A12A1LO/synthesizer boardE4440-60283
A12A2Sampling oscillator boardE4440-60009
A13Front end driver boardE4446-60166Serial prefix 4251 and above
A13Front end driver replacement kitE4446-60014Serial prefix 4222 and below
A14Input attenuator, 10 dB / switch33326-60009
A15Input attenuator, 60 dB33325-60009
A18YTO, 2.9 to 7 GHzE4440-60023
A19SBTX/RYTHM assembly, 44 GHz5087-7184Refurbished 5087-6184
A20Low band assemblyE4440-60217
A21FELOMA (frequency extended LO multiplying amplifier)5087-7126
A22Preamp assembly, 3 GHzE4440-60444Option 1DS (firmware A.06.04 or later)
A23Floppy disk drive0950-2782
A25MotherboardE4440-60001
A26CPU (processor)E4406-60060Without DRAM / flash
A26A1128 MB DRAM assemblyE4406-60053
A26A264 MB flash memory replacement kitE4440-60094
A26BT1Lithium battery, 3 V1420-0314
A27Electronic attenuatorE4440-60456Option B7J
A28Audio out boardE4406-60025
A29SBTX driver boardE4446-60005
A30FIFA first IF amplifierE4446-60007
A34Unpreselected mm-wave mixer1NB7-8539Option 123
A35Mixer bias boardE4440-60254Option 123
A36Millimeter preamp, 50 GHz0955-1617Option 110
A37Audio digitizerE4440-60264Option 107
A38Option driver assemblyE4440-60459Option 110/123
A39USB/memory boardE4440-60252Option 111, 115, 117
B1-B3FanE4406-60159
FL1Low pass filter, 3 GHz0955-0988
FL2Band pass filter, 3.9214 GHz0955-1074

Part obsolete? Many E4447A assemblies are no longer sold new. Analyzer Repair Inc. repairs E4447A RF assemblies, YTOs, IF and synthesizer boards at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

From the Agilent PSA Service Guide (E4440-90616).