Keysight (Agilent HP) E5061A Repair Service
Product model
E5061A ENA Series Network AnalyzerServices
Description

Repair prices
E5061A Specifications – click to expand
E5061A Specifications
- ENA-series vector network analyzer
- Windows-based with color display, LAN, USB and GPIB interfaces
E5061A Options – click to expand
E5061A Options
- 150/250, 50 Ohm Transmission/Reflection or S-Parameter Test Set. 175/275, 75 Ohm Transmission/Reflection or S-Parameter Test Set
- 1E5, High-Stability Frequency Reference
- 100, Fault Location and Structural Return Loss
Agilent / HP E5061A Error Codes – click to expand
Agilent E5061A Error Messages
Error and fault messages for the Agilent E5061A network analyzer (1.5 GHz), with the assemblies the service manual points to. Errors show on a red bar at the lower left; numbered ones can be read over GPIB.
E5061A notes
- The E5061A runs a power-on self test at every start; a failure is reported as error 241 and broken down by the internal test in the service menu.
- If the rear fan does not run, the A50 power supply is suspect; check the +3.3 V / +5 V monitor LEDs on the A21 PCI DSP card.
E5061A hardware errors
| Error | Message | Meaning |
|---|---|---|
| 220 | Phase lock loop unlocked | A synthesizer PLL lost lock during measurement. The service manual points to the A1 source board. |
| 241 | Power on test failed | The power-on self test found a hardware fault. Run the internal test from the service menu; the first failed test identifies the board (A1 source, A2 receiver board). |
| 240 | Calibration data lost | The system calibration file is missing or damaged at startup. The analyzer needs its adjustment data restored or regenerated. |
| 221 | Port 1 receiver overload | Too much power into port 1. If it appears with nothing connected, the analyzer is faulty. |
| 222 | Port 2 receiver overload | Same as 221 for port 2. |
| 61 | Power unleveled | Requested output power is outside the leveling range after correction. |
| — | Fatal Error: Failed to Initialize DSP / DSP Driver | The A21 PCI DSP card or A20 digital motherboard did not start. Also shown after a failed firmware update. |
| — | Fatal Error: Failed to Update DSP Code | Writing DSP firmware to flash on the A21 PCI DSP card failed; reinstall firmware, then suspect A21 PCI DSP card or A20. |
E5061A calibration, file and setup errors
| Error | Message |
|---|---|
| 20 | Additional standard needed |
| 22 | Calibration method not selected |
| 30 | Valid ECal module not found |
| 31 | Failed to configure ECal module |
| 41 | Peak not found |
| 53 | Log sweep requires 2 octave minimum span |
| 100 | Failed to read file |
| 101 | Failed to write file |
| 105 | Recall failed |
| 106 | Save failed |
| 120 | Printer error |
| 200 | Option not installed |
SCPI error ranges
| Range | Class |
|---|---|
| −100 to −199 | Command errors |
| −200 to −299 | Execution errors |
| −300 to −399 | Device-specific errors |
| −400 to −499 | Query errors |
Hardware error on your E5061A? Analyzer Repair Inc. repairs and calibrates the Agilent E5061A in-house in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.
Based on the Agilent E5061A/E5062A ENA Service Manual (E5061-90140).
Agilent / HP E5061A Replaceable Parts – click to expand
Agilent E5061A Replaceable Assemblies and Part Numbers
Major replaceable assemblies for the Agilent E5061A, from the Agilent E5061A/E5062A ENA Service Manual (E5061-90140).
| Reference | Assembly | Agilent part number | Notes |
|---|---|---|---|
| A1 | Source board | E5062-62111 | Exchange E5062-69111 |
| A2 | Receiver, 50 Ω | E5062-62011 | Options 150, 250; exchange E5062-69011 |
| A2 | Receiver, 75 Ω | E5062-62012 | Options 175, 275; exchange E5062-69012 |
| A20 | Digital motherboard | E5062-62003 | |
| A21 | PCI DSP card | E5062-66521 | |
| A24 | PCI GPIB card | E5062-65000 | |
| A22 | Front panel keyboard | E5062-66522 | |
| A27 | Hard disk drive | E5062-83102 | |
| A28 | Floppy disk drive | 0950-5083 | |
| A50 | Power supply | 0950-4773 | |
| A52 | Inverter | 0950-4091 | |
| — | Analog motherboard | E5062-66510 | |
| — | LCD assembly | E5070-60101 | Touch-panel LCD E5070-60102 with Option 016 |
Part obsolete? Many E5061A assemblies are no longer sold new. Analyzer Repair Inc. repairs E5061A boards and RF modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.
From the Agilent E5061A/E5062A ENA Service Manual (E5061-90140).