Keysight (Agilent HP) E8361C Repair Service
Product model
E8361C PNA Network Analyzer 10 MHz to 67 GHzServices
Description

Repair prices
E8361C Specifications – click to expand
E8361C Specifications
94 dB of dynamic range and
E8361C Options – click to expand
E8361C Options
- Time domain for 6 GHz model N52310-010
- Time domain for 13.5, 20, 40 or 50 GHz model N52300-010
- Frequency-offset measurements N52300-080
- Scalar-calibrated converter measurements
- (Option 080 required)
- N52300-082
- 4-port measurement application N52300-550
- N-port capabilities for 6, 13.5 or 20 GHz model N52310-551
- N-port capabilities for 40 or 50 GHz model N52300-551
Agilent / HP E8361C Error Codes – click to expand
Agilent E8361C Error Messages
Error and fault messages for the Agilent E8361C network analyzer (67 GHz), with the assemblies the service manual points to.
E8361C on-screen error messages
| Message | Meaning and where to look |
|---|---|
| PHASE LOCK LOST | The A11 phase lock board sees too little reference-receiver signal to lock. Cause can be in the source, signal separation (couplers, switch) or receiver path. Check the A10 frequency reference output and the reference receiver first. |
| SOURCE UNLEVELED | On the E8361C check W38/W39 from the A23/A24 SOMA 70s to A16 and W40 to the A21 SOMA 50. One port only: replace that SOMA 70. Both ports: replace the A16 test set motherboard. |
E8361C failed performance test – likely cause
| Failed test | What it points to |
|---|---|
| Source power accuracy / linearity | Run the source calibration adjustment; if it still fails, the source (A12 source) is suspect. |
| Source maximum power output | Follow the signal through the separation path (switch, couplers, attenuators) to find the loss. |
| Frequency accuracy | Measure the rear-panel 10 MHz REF OUT; more than about 10 Hz off calls for the 10 MHz reference adjustment. |
| Trace noise | Rarely fails alone; run the receiver calibration adjustment, then suspect the A6 SPAM board. |
| Receiver compression / noise floor / dynamic accuracy | Points to the A31–A34 receivers or the A6 SPAM board. Receivers are identical, so swapping two channels confirms a bad one. |
| Calibration coefficients | Tracking terms only: run the receiver calibration adjustment. Any other error term failing means a hardware fault. |
| System crosstalk | Loose RF cable, or isolation loss in the receivers or the receiver/test set motherboard. Tighten all cables and receiver screws first. |
SCPI error ranges
| Range | Class |
|---|---|
| −100 to −199 | Command errors |
| −200 to −299 | Execution errors |
| −300 to −399 | Device-specific errors |
| −400 to −499 | Query errors |
Hardware error on your E8361C? Analyzer Repair Inc. repairs and calibrates the Agilent E8361C in-house in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.
Based on the Agilent E8361C PNA Service Guide (E8361-90008).
Agilent / HP E8361C Replaceable Parts – click to expand
Agilent E8361C Replaceable Assemblies and Part Numbers
Major replaceable assemblies for the Agilent E8361C, from the Agilent E8361C PNA Service Guide (E8361-90008).
| Reference | Assembly | Agilent part number | Notes |
|---|---|---|---|
| A1 | Keypad assembly | N5230-40002 | |
| A2 | Display assembly | N5230-60045 | |
| A3 | Front panel interface board | N5230-60132 | |
| A4 | Power supply assembly | 0950-4599 | |
| A6 | SPAM board | E8356-60006 | |
| A8 | Fractional-N synthesizer board | E8364-60099 | A9 second Frac-N with Option 080 |
| A10 | Frequency reference board | E8364-60136 | |
| A11 | Phase lock board | E8364-60137 | |
| A12 | Source 20 assembly | E8361-60002 | Exchange E8361-69002 |
| A13 | Frequency offset receiver board | E8364-60097 | Option 080 |
| A14 | System motherboard | E8364-60100 | |
| A15 | CPU board, 1.1 GHz | E8364-60204 | |
| A16 | Test set motherboard | E8364-60138 | |
| A17 | LOMA 12 | 5087-7247 | Exchange 5087-6247 |
| A18 | MA 24 | 5087-7248 | Exchange 5087-6248 |
| A19 | Splitter | 5087-7139 | |
| A20 | LODA | 5087-7777 | Exchange 5087-6777 (was 5087-7129) |
| A21 | SOMA 50 | 5087-7130 | Exchange 5087-6130 |
| A22 | Switch | 5087-7238 | Exchange 5087-6238 |
| A23, A24 | SOMA 70 | 5087-7249 | 2 required |
| A25, A26 | Test port couplers | 5087-7225 | |
| A27–A30 | First converters (mixers) | 5087-7258 | Exchange 5087-6258; 4 required |
| A31–A34 | Receiver modules | E8364-60134 | 4 required |
| A35 | Receiver motherboard | E8364-60135 | |
| A36, A37 | 50 dB source step attenuators | 84905-60001 | Option UNL |
| A38, A39 | Bias tees | 5086-7020 | Option UNL |
| A45 | 70 GHz reference switch | 5087-7271 | |
| A49 | IF multiplexer board | E8361-60043 | |
| A41 | Programmed hard disk drive | N5230-60048 | Parallel ATA |
| B1 | Fan | 3160-1085 | 3 required |
Part obsolete? Many E8361C assemblies are no longer sold new. Analyzer Repair Inc. repairs E8361C boards and RF modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.
From the Agilent E8361C PNA Service Guide (E8361-90008).