Keysight (Agilent HP) N9030A Repair Service

N9030A PXA Signal Analyzer, 3 Hz to 50 GHz
Our technicians repair and calibrate the Keysight Agilent N9030A Signal Analyzer in-house. Analyzers are our business. Free evaluations, veteran owned.
Designed for the real world, Keysight’s future-ready PXA signal analyzer is the evolutionary replacement for your current highperformance analyzer. It helps you sustain past achievements, enhance current designs, and accelerate future innovations. Take control with the PXA signal analyzer—and drive your evolution.
n9030a-pxa-analyzer
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
N9030A Specifications – click to expand

N9030A Specifications

  • Frequency ranges
  • Up to 325 GHz and beyond with external mixing
  • Analysis bandwidth 10, 25, 40, and 160 MHz
  • Displayed average noise level (DANL) –172 dBm at 1 GHz, preamplifier and noise floor extension on
  • Third-order intermodulation (TOI) distortion +22 dBm at 1 GHz W-CDMA ACLR dynamic rangern–83 dBc (–88 dBc nominal) with noise correction on Phase noise –132 dBc/Hz at 10 kHz offset (1 GHz carrier) Amplitude accuracy ± 0.19 dB Real-time bandwidth 160 MHz; up to 50 GHz frequency range
  • Probability of intercept 100% with signal durations as short as 3.57 µs
SpecificationValue
Minimum3 Hz - Maximum: 3.6, 8.4, 13.6, 26.5, 43, 44, 50 GHz
N9030A Options – click to expand

N9030A Options

  • Frequency range options: 503, 508, 513, 526, 543, 544, 550 (e.g. 526 = 26.5 GHz)
  • P03-P50, Preamplifier
  • EA3, Electronic Attenuator
  • B25, 25 MHz Analysis Bandwidth
  • PFR, Precision Frequency Reference
Agilent / HP N9030A Error Codes – click to expand

Agilent N9030A Error Messages

Condition (status) errors for the Keysight / Agilent N9030A PXA signal analyzer (3.6 to 50 GHz depending on option). Each error has a "cleared" twin numbered +1000 (for example 509 / 1509).

N9030A notes

  • Options 503–526 use the 0–4 dB and 0–66 dB attenuator pair; 543/544/550 use the 0–10 dB and 0–60 dB pair and their own front end and preselector.
  • Many N9030A numbers replace older PSA numbers (for example Align Now, All was 64 on the PSA and is still 64 here).

N9030A hardware and alignment errors

ErrorMessageMeaning
40TG Alignment FailureTracking generator alignment failed.
42RF Alignment FailureRF alignment failed: check the A13 front end and the internal 50 MHz calibrator path.
44IF Alignment FailureIF alignment failed: A2 analog IF or A3 digital IF.
46LO Alignment FailureLO alignment failed: A14 synthesizer or A20 YTO.
48ADC Alignment FailureADC alignment failed: A3 digital IF.
60Align Now, RF requiredRun Align Now, RF.
64Align Now, All requiredRun Align Now, All. Recurring after a good alignment means a hardware fault.
141Input OverloadToo much signal at the IF/ADC. Add attenuation; with no input applied it points to the A13 front end or A2 analog IF.
403Source UnleveledTracking source cannot hold level.
419Preselector OverloadOverload at the A12 YTF preselector path.
501Source Synth UnlockedTracking source synthesizer unlocked.
503Frequency Reference UnlockedReference loop unlocked: A16 reference assembly or the external reference.
5052nd LO Unlocked2nd LO unlocked: A13 front end / A16 reference.
509LO Unlocked1st LO unlocked: A14 synthesizer, A20 YTO, A15 front end controller.
513IF Synthesizer UnlockedIF synthesizer unlocked: A16 reference / A3 digital IF.
515Calibration Oscillator UnlockedInternal calibrator unlocked: A16 reference.
521External ref missing or out of rangeExternal reference signal missing, too low, or not at the entered frequency.
301Meas UncalSweep too fast for the RBW; press Auto Couple.
6Align RF SkippedRF alignment skipped because a 50 MHz signal is present at the input (warning).

SCPI error ranges

RangeClass
−100 to −199Command errors
−200 to −299Execution errors
−300 to −399Device-specific errors
−400 to −499Query errors

Hardware error on your N9030A? Analyzer Repair Inc. repairs and calibrates the Agilent N9030A in-house in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

Based on the Keysight N9030A PXA Service Guide (9018-04011).

Agilent / HP N9030A Replaceable Parts – click to expand

Agilent N9030A Replaceable Assemblies and Part Numbers

Board and RF module part numbers for the Agilent N9030A; exchange numbers are factory-rebuilt units.

ReferenceAssemblyAgilent part numberNotes
A1A2Front panel interface boardW1312-60137Serial > MY49431900 / SG49430063 / US51160001 (earlier W1312-63092)
A1A3LCD, LED backlight2090-0997
A1A4DC-DC converter (LED backlight)0950-5235
A1A5Front panel daughter boardN9020-63193
A2Analog IF assemblyN9020-60254
A3Digital IF assemblyN9020-60016
A4CPU boardN9020-60247Serial prefix 5533 and above (N9020-60135 from 5136/US5135; N9020-60081 earlier)
A5Solid state drive, Windows 7N9020-60202
A6Power supply0950-4934
A7Midplane assemblyW1312-63095
A8MotherboardN9030-63002
A9RF attenuator A, 0–4 dB33360-60008Options 503–526
A9RF attenuator A, 0–10 dB33326-60013Options 543/544/550
A10RF attenuator B, 0–66 dB33321-60083Options 503–526
A10RF attenuator B, 0–60 dB33325-60020Options 543/544/550
A11Low band switchE4410-60121Options 503–526
A11Low band switchN9020-60051Options 543/544/550
A12YTF preselectorN9020-60229Options 503–526
A12YTF preselectorN9030-60024Options 543/544/550
A13Front end assemblyN9020-60143Options 503–526, serial prefix 5235 and above (N9020-60173 earlier)
A13Front end assemblyN9020-60144Options 543/544/550, serial prefix 5235 and above (N9030-60010 earlier)
A14Synthesizer assemblyN9020-60008
A15Front end controllerN9020-60172
A16Reference assemblyN9020-60188
A20YTO assemblyN9020-60009
A25Wideband analog IFN9020-60044Option B1X
A26Wideband digital IFN9020-60257Option B1X
A18, A19BBIQ interface / main boardN9020-60094Option BBA (main board N9020-60093)
J1Type-N input connectorN9039-60030Options 503–526 (3.5 mm N9020-60196 with C35; 2.4 mm N9030-60011 for 543/544/550)
B1–B3Fan3160-4199

Part obsolete? Many N9030A assemblies are no longer sold new. Analyzer Repair Inc. repairs N9030A boards and RF modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

From the Keysight N9030A PXA Service Guide (9018-04011).

Located in: Network Analyzers