• no items with letter:
    A
  • no items with letter:
    B
  • no items with letter:
    C
  • no items with letter:
    D
  • show items with letter:
    E
  • show items with letter:
    F
  • no items with letter:
    G
  • no items with letter:
    H
  • no items with letter:
    I
  • no items with letter:
    J
  • no items with letter:
    K
  • show items with letter:
    L
  • no items with letter:
    M
  • show items with letter:
    N
  • no items with letter:
    O
  • no items with letter:
    P
  • no items with letter:
    Q
  • show items with letter:
    R
  • no items with letter:
    S
  • show items with letter:
    T
  • show items with letter:
    U
  • no items with letter:
    V
  • no items with letter:
    W
  • no items with letter:
    X
  • no items with letter:
    Y
  • show items with letter:
    Z
  • show items with letter:
    0-9

Keysight (Agilent HP) 8562E Repair Service

8562E spectrum analyzer 30Hz-13.2GHz
Complete Agilent HP 8562E Spectrum Analyzer repair and calibration, even after factory support ends. Free estimates and quick turnaround times.

We are specialists in the repair of the Agilent/HP 8562E spectrum analyzers.The 8562E has a frequency range of 30 Hz to 13.2 GHz, which covers the spur-search ranges specified by leading standards organization in Europe and the United States for GSM. With its fast DLP execution speed, Agilent 8562E reduces your test times in production.

8562e
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
8562E Specifications – click to expand

8562E Specifications

SpecificationValue
Frequency accuracy± 103 Hz (CF = 1 GHz)
Frequency Span Range0 Hz, 100 Hz to maximum frequency
Frequency Readout Accuracy Span >2 MHz x N± (freq. readout x freq. ref. accuracy +5% x span +15% x RBW +10 Hz)
Frequency Readout Accuracy Span =2 MHz x N± (freq. readout x freq. ref. accuracy +1% x span +15% x RBW +10 Hz)
Sweep Time RangeSpan = 0 Hz: 50 µs to 6,000 s; Span =100 Hz: 50 ms to 100 ks
Resolution bandwidth range1 Hz to 2 MHz
Resolution bandwidth Accuracy1 Hz to 300 kHz: ±10%; 1 MHz: ±25%; 2 MHz: +50%, –25%
Video Bandwidth Range1 Hz to 3 MHz in a 1, 3, 10 sequence
Amplitude RangeDisplayed average noise level to +30 dBm
Average noise level (narrowest RBW)-151 dBm
Optimum dynamic range 2nd/3rd order95 dB / 108 dB
Amplitude accuracy± 2.1dB
8562E Options – click to expand

8562E Options

  • 001, Second IF Output (310.7 MHz)
  • 005, Alternate Sweep Out (cannot be used with Option 002)
  • 007, Digitized Fast Time-Domain Sweeps
  • 008, Signal Identification
  • 042 Protective Soft Carrying Case/Backpack
  • 0Q8, Service Training
  • 103, Delete Precision Frequency Reference, and 1 Hz and 3 Hz RBWs
  • 104, Delete Mass Memory Module
HP / Agilent 8562E Error Codes (ERR 100–922) – click to expand

HP / Agilent 8562E Error Codes

Every fault the HP 8562E (30 Hz to 13.2 GHz) detects shows up as an ERR number and short message at the lower right of the display. To see every stored error, press RECALL > MORE 1 OF 2 > RECALL ERRORS and scroll with the arrow keys. One fault often triggers several codes, so the lowest hardware code is the best place to start.

First step: save your state, press PRESET, CAL, REALIGN LO & IF and let it finish. If a code between 200 and 799 returns, the 8562E needs service.

8562E notes

  • The 8562E has a CRT display driven by a high-voltage module. A dim, blank or distorted screen usually has no ERR code at all.
  • FADC codes ERR 202–207 and ERR 760 only apply to an 8562E with Option 007 (fast digitized sweeps).
  • The 8562E input can be AC or DC coupled (AC is the power-on default). Keep DC under 0.2 V when DC coupled to protect the input mixer.
  • Above 2.9 GHz the 8562E uses a YIG preselector. Bad stored preselector peak data shows as ERR 704; use FACTORY PRSEL PK to restore it.
  • Using the 85629B test and adjustment module on the 8562E needs the 562E/TAM interface software.

Most common 8562E hardware errors

CodeDisplayMeaning
ERR 300 / 301YTO UNLKThe YTO (first LO) loop will not lock.
ERR 334LO AMPLFirst LO amplifier not leveled. Put a 50 ohm load on 1ST LO OUTPUT before suspecting the YTO.
ERR 337FN UNLKFractional-N loop cannot lock.
ERR 317 / 318FREQ ACCYTO main-coil tuning DAC at its limit.
ERR 319 / 320WARN COA / FINYTO tune DAC near its limit (warning).
ERR 351 / 353SETL FLD / SRCH FLDYTO error voltage won't settle, or no usable YTO DAC value found.
ERR 499CAL UNLKIF cal oscillator unlocked.
ERR 700EEROMStored calibration data (EEROM A2U501) checksum error.
ERR 718BATTERY?Memory battery low; stored states are being lost.
ERR 335SMP UNLKSampling oscillator loop unlocked.
ERR 333600 UNLK600 MHz reference loop unlocked.
ERR 704PRESELCTPreselector peak data checksum error or wrong-mode recall.
ERR 906OVENCOLDOCXO not warmed up. Normal right after power-on.

Code ranges

RangeArea
100–199Programming / GPIB (not a hardware fault)
200–299ADC
300–399LO and RF
400–599IF (automatic IF alignment)
600–699System
700–799Digital / memory
800–899Option modules
900–999User / setup

Complete 8562E error code list

100–199 Programming errors – 8562E

CodeDisplayMeaning
ERR 100NO PWRONSaved power-on state is bad; default state loaded instead.
ERR 101NO STATEThe state you tried to recall was never saved or is invalid.
ERR 102–105, 111# ARGMTSCommand is missing arguments.
ERR 106ABORTED!Operation aborted; GPIB parser was reset.
ERR 107HELLO ??No GPIB listener found.
ERR 108TIME OUTTimed out while acting as bus controller.
ERR 109CtrlFailCould not take control of the GPIB bus.
ERR 110NOT CTRLAnalyzer is not the system controller.
ERR 112??CMD??Command not recognized.
ERR 113 / 114 / 115FREQ NO! / TIME NO! / AMPL NO!Command does not accept frequency, time or amplitude units.
ERR 116?UNITS??Units not recognized.
ERR 117–122NOP NUM / EP / UPDN / ONOF / ARG / TRCArgument type not allowed for this command (numeric, enable, UP/DN, ON/OFF, AUTO/MAN or trace register).
ERR 123 / 124NOP ABLK / NOP IBLKA-block or I-block data format not allowed here.
ERR 125NOP STRNGStrings not allowed for this command.
ERR 126NO ?Command cannot be queried.
ERR 127BAD DTMDInvalid detector mode.
ERR 128PK WHAT?Invalid peak search parameter.
ERR 129PRE TERMA-block data ended early.
ERR 130BAD TDFArgument only valid with the TDF command.
ERR 131–138, 140?? AM/FM, !FAV/RMP, !INT/EXT, ??? ZERO/CURR/FULL/LAST, !GRT/DSP, ?? PWRONThat keyword is not a valid argument for the command sent.
ERR 139PLOTONLYArgument only valid with the PLOT command.
ERR 141 / 142BAD ARGFDIAG argument or query problem.
ERR 143NO PRESLCommand needs preselector hardware.
ERR 144COUPL??Bad COUPLING argument; use AC or DC. The 8562E supports both.

200–299 ADC errors – 8562E

CodeDisplayMeaning
ERR 200 / 201SYSTEMADC driver or controller interaction fault; look for other codes.
ERR 202–207FADC CALFast ADC offset or slope calibration failed (linear, log, or log 2x expand). Only on an 8562E with Option 007.
ERR 250OUTOF RGSignal at the ADC is out of range.
ERR 251NO IRQProcessor is not getting interrupts from the ADC.

300–399 LO and RF errors – 8562E

CodeDisplayMeaning
ERR 300 / 301YTO UNLKYTO (first LO) phase-lock loop unlocked.
ERR 302 / 327OFF UNLKOffset roller oscillator unlocked.
ERR 303XFR UNLKTransfer roller oscillator unlocked.
ERR 304ROL UNLKMain roller oscillator unlocked.
ERR 305, 306, 312, 324, 325FREQ ACCCoarse or fine adjust DAC cannot null the MAINSENSE voltage.
ERR 307FREQ ACCTransfer oscillator pretune DAC out of range.
ERR 308 / 309FREQ ACCOffset oscillator pretune DAC outside limits at the low or high end.
ERR 310 / 311 / 322FREQ ACCMain roller pretune DAC ran out of range during tuning.
ERR 313FREQ ACCSampler and roller oscillator frequencies don't add up to the required YTO start frequency.
ERR 314 / 315FREQ ACCSpan calibration problem: unlocked main roller, missing sweep ramp, or span attenuator DAC out of range.
ERR 316 / 321FREQ ACCMain roller tuning sensitivity measured as zero or less.
ERR 317 / 318FREQ ACCYTO main-coil coarse or fine DAC at its limit.
ERR 319 / 320WARN COA / WARN FINYTO coarse or fine tune DAC near its limit (warning).
ERR 326FREQ ACCFine adjust DAC near the end of its range.
ERR 328 / 329FREQ ACCRoller fine or coarse DAC sensitivity measured as zero or less.
ERR 331FREQ ACCInvalid YTO frequency.
ERR 333600 UNLK600 MHz reference loop unlocked.
ERR 334LO AMPLFirst LO distribution amp not leveled; 1ST LO OUTPUT must be terminated in 50 ohms.
ERR 335SMP UNLKSampling oscillator unlocked.
ERR 33610 MHzCal oscillator would not lock on the internal 10 MHz reference.
ERR 337FN UNLKFractional-N loop cannot lock.
ERR 338 / 339 / 340LOMA AGC / SBTX AGC / SAMP AGCLO multiplier/amplifier main, SBTX or sampler loop AGC not leveled.
ERR 351SETL FLDYTO error voltage not settling.
ERR 352TWID FLDError-voltage DACs cannot reach their quiescent point.
ERR 353SRCH FLDNo acceptable YTO DAC value found.
ERR 354LK ITERSLock routine stopped without locking.
ERR 355SMP CALSampler unlocked during calibration; stays until a good recalibration.
ERR 356–361SPAC CALSpan accuracy calibration sweep failed (bucket search, intersection, start or percent-of-span correction out of range).

400–599 IF errors – 8562E

These come from the automatic IF alignment. It aligns amplitude first, then the resolution bandwidths in this order: 300 kHz, 1 MHz, 2 MHz, 100 kHz, 30 kHz, 10 kHz, 3 kHz, 1 kHz, 300 Hz down to 1 Hz. Bandwidths aligned before the failure are usually still good; ones after it are suspect.

CodeDisplayMeaning
ERR 400–404AMPL <300 / 300 / 1K / 3K / 10KCannot set the amplitude of that resolution bandwidth.
ERR 405–449RBW 300 / 1K / 3K / 10KCannot align an individual pole of a crystal resolution bandwidth filter.
ERR 450–452IF SYSTMGeneral IF hardware failure; look for other codes.
ERR 454–465, 571AMPLCannot adjust the step gain amplifiers.
ERR 466LIN AMPLCannot adjust the linear amplitude scale.
ERR 467–470, 552–570, 573–577, 591, 592LOG AMPLCannot adjust amplitude in log scale (log amplifier).
ERR 471–478, 583–590RBW 30K / 100K / 300K / 1MCannot align that LC resolution bandwidth.
ERR 483–498, 517–521RBW 100 Hz to 10KCannot align that narrow resolution bandwidth.
ERR 499CAL UNLKIF cal oscillator unlocked.
ERR 500–507, 572AMPL 30K / .1M / .3M / 1MCannot set amplitude of an LC resolution bandwidth.
ERR 508–511AMPLNot enough gain during LC bandwidth calibration.
ERR 512–516RBWNot enough gain during crystal bandwidth calibration.
ERR 522–525RBW 10KCannot align symmetry of a 10 kHz filter pole.
ERR 526–540RBW <300Digital narrow-bandwidth (under 300 Hz) calibration failed: sampling timeout, step gain, DC level, demod data, VCXO, flatness, gain, or cal oscillator count/lock problems.
ERR 550IDCALOSCCal oscillator ID doesn't match; incompatible hardware.
ERR 551ID LOGBDLog board ID doesn't match; incompatible hardware.
ERR 578–580, 593–599LOG AMPL / LOG TUNE / LOG OFST / LOG ATTN / LOG FID / LOG GAINDC logger calibration error (limiter, attenuator level, fidelity offset, gain or tune).
ERR 581 / 582AMPLCannot adjust 100 kHz and the 10 kHz-and-below bandwidths; 582 adds a bad cal oscillator sweep-rate calibration.

600–699 System errors – 8562E

CodeDisplayMeaning
ERR 600 / 601SYSTEMHardware/firmware interaction fault; look for other codes.
ERR 650OUTOF RGSignal at the ADC is out of range.
ERR 651NO IRQProcessor is not getting interrupts from the ADC.

700–799 Digital and memory errors – 8562E

CodeDisplayMeaning
ERR 700EEROMChecksum error in EEROM A2U501 (stored calibration data).
ERR 701 / 703AMPL CALFrequency response correction data checksum error.
ERR 702ELAP TIMElapsed-time data checksum error.
ERR 704PRESELCTPreselector peak data checksum error, or internal/external preselector data recalled in the wrong mode.
ERR 705–710ROM U306–U311Program ROM checksum error (A2U306 through A2U311).
ERR 711–716RAM U300–U305System RAM checksum error (A2U300 through A2U305).
ERR 717BAD uP!!Microprocessor not fully working.
ERR 718BATTERY?Nonvolatile RAM not holding; check the memory battery.
ERR 719 / 720MODEL #? / OPTION #?Cannot identify the model or an installed option.
ERR 721AMPC RAMAmplitude correction (ampcor) data checksum error.
ERR 750–759SYSTEMFirmware math or system fault (divide by zero, overflow, underflow, unknown); look for other codes.
ERR 760NO FADCFast ADC failed its power-on check. Only on an 8562E with Option 007.

800–899 Option module errors – 8562E

Reported by an attached rear-panel module such as the 85620A mass memory module or the 85629B test and adjustment module. See that module's manual.

900–999 User and setup errors – 8562E

CodeDisplayMeaning
ERR 900TG UNLVLTracking generator output not leveled (only with Option 002 tracking generator models).
ERR 901TGFrqLmtTracking generator unleveled because the start frequency is below its 300 kHz limit.
ERR 902BAD NORMCurrent state doesn't match the state stored for normalization.
ERR 903 / 904A > DLMT / B > DLMTTrace A or calibration trace B is off screen with trace math or normalization on.
ERR 905EXT REFCannot lock to the external reference; check that it is 10 MHz and in tolerance.
ERR 906OVENCOLDOCXO oven still cold.
ERR 907DO IFCALIF calibration still running, or still needed after an OVENCOLD condition.
ERR 908–911BW>>SPCG / SPAN<ACP / SPAN>ACP / ACP STATEAdjacent channel power setup is invalid (bandwidth, spacing or span).
ERR 919 / 920SP <CHBW / RBW>CHBWChannel power setup invalid: spacing too narrow or resolution bandwidth too wide.
ERR 921 / 922AMPCORAmpcor corrected data that was above or below the screen, so its accuracy is unspecified.

8562E still showing hardware errors after a realign? Our lab in Cocoa, Florida has been repairing 8560 series analyzers for years. Free estimate once the unit is here; you approve before any work starts. Contact us for service.

Code list based on the Agilent 8560 E-Series and EC-Series User's Guide, Chapter 9.

HP / Agilent 8562E Replaceable Parts – click to expand

HP 8562E Replaceable Assemblies and Part Numbers

Assembly-level HP / Agilent part numbers for the 8562E, from the HP 8562E Service Guide (08562-90216). Where more than one number is listed, the part depends on options or serial prefix (shown in the Notes column).

ReferenceAssemblyHP part numberNotes
A1A1Keyboard board08562-60140
A1A2RPG (rotary knob) assembly0960-0745
A2Controller assembly08564-60010
A3Interface assembly08563-60098
A4Log amplifier / cal oscillator assembly08563-60076
A5IF filter assembly08563-60023
A6Power supply assembly (includes A6A2)08564-60008
A6A1High voltage assembly5062-7089
A6A2Regulator board assembly08564-60009
A7Switched LO distribution amplifier5086-7885
A7Switched LO distribution amplifier (rebuilt exchange)5086-6885
A8Low band mixer5086-7982
A9Port attenuator5086-7822
A10YIG-tuned filter/mixer (RYTHM)5086-7981
A10RYTHM (rebuilt exchange)5086-6981
A11Leveled YTO5086-7906
A11Leveled YTO (rebuilt exchange)5086-6906
A13Second converter5086-7959
A14Frequency control assembly08563-60090
A15RF assembly (standard)08563-60091
A15RF assembly (TCXO)08563-60092Option 103
A15RF assembly (SIG ID)08563-60093Option 008
A15U100Sampler5086-7806
A16Fast ADC assembly08563-60030Option 007
A17CRT driver assembly08563-60101
A18MP1CRT wiring assembly5062-7095
A18V1CRT tube, 6.7 in2090-0225
A19HP-IB assembly08562-60042
A20Battery assembly5062-7755
BT1Battery, 3.0 V lithium1420-0341
A21OCXO, 10 MHz5063-0245Standard
B1Fan assembly5061-9036
FL1Low-pass filter0955-0703
FL2Low-pass filter, 4.4 GHz0955-0519
FL4Line filter assembly5061-9032
F1Fuse 5 A 250 V (230 VAC)2110-0709
F1Fuse 5 A 125 V (115 VAC)2110-0756

Boards marked for HP's rebuilt exchange program use the same number with the 7th digit changed to 9 (for example 08563-60044 becomes 08563-69044).

Part obsolete? Most 8562E assemblies are no longer sold new. Analyzer Repair Inc. repairs 8562E boards and RF modules at the component level in Cocoa, Florida. Free estimate once the unit is in our lab. Contact us for service.

From the HP 8562E Service Guide (08562-90216).